Predictors of Variability in Coronary Flow Reserve and Microvascular Resistance Measurement Using the Bolus Thermodilution Technique.

Galo Jason, Maksoud Aziz, Shah Samit, Toleva Olga, Kobayashi Yuhei, Fahed Akl C, Shlofmitz Evan, Sharma Alok, Buijs Jorn Op den, Spinelli Joia, Rapoza Richard, Bergmark Brian, Hashim Hayder

СтатьяМногоцентровое исследованиеНаблюдательное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Bolus thermodilution is the most commonly used invasive assessment for coronary microvascular dysfunction (CMD). However, variability in thermodilution transit time measurements (TTM) affects calculated Coronary Flow Reserve (CFR) and Index of Microvascular Resistance (IMR) values. This study determined variables associated with transit time variability and quantified confidence intervals for the range of measured CFR and IMR values.

METHODS

The CoroFlow Cardiovascular System (Coroventis, Sweden) and PressureWire™ X Guidewire (Abbott) were used to collect 2541 TTM from 259 vessels at 14 US sites included in the FlowLab Study. Correlations with baseline patient and procedural variables were determined by both linear and mixed linear models. Mixed linear models with the physician or vessel as the random effect were computed. Monte Carlo simulations of CFR and IMR were performed to determine 95% confidence intervals for thresholds of 2.5 for CFR and 25 for IMR.

RESULTS

Mixed linear analysis with vessel as the random effect provided the best model fit (R2 = 0.948). In this model, RCA vessel location (p < 0.001) and pressure sensor distance < 5 cm (p < 0.005) were significantly associated with transit time. Monte Carlo simulations identified a "gray zone" range of 2.1-3.1 for CFR and 19-31 for IMR.

CONCLUSIONS

Minimizing variability in invasive CMD assessment is essential for improving diagnostic accuracy and guiding clinical decision-making as testing becomes more widespread. In this study, target vessel choice (RCA) and pressure sensor placement were identified as key modifiable sources of transit time variability. Clinically applicable confidence intervals around CFR and IMR values were defined.

Для цитирования: Galo Jason, Maksoud Aziz, Shah Samit, Toleva Olga, Kobayashi Yuhei, Fahed Akl C, Shlofmitz Evan, Sharma Alok, Buijs Jorn Op den, Spinelli Joia, Rapoza Richard, Bergmark Brian, Hashim Hayder. Predictors of Variability in Coronary Flow Reserve and Microvascular Resistance Measurement Using the Bolus Thermodilution Technique.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2025. DOI: 10.1002/ccd.70301.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.