BACKGROUND

The thermodilution technique, employing bolus injections to measure coronary flow reserve (CFR) and the index of microcirculatory resistance (IMR), is widely used to assess coronary microvascular function. However, the potential influence of catheter size and operator differences on these measurements has not been systematically evaluated.

AIMS: This study aimed to evaluate the feasibility and consistency of thermodilution-based CMD assessment with respect to different catheter sizes and operators.

METHODS

A total of 320 coronary arteries were assessed using a 0.014-inch pressure-temperature sensor-tipped guidewire. The data set included both angiographically normal vessels and those with varying degrees of stenosis.

RESULTS

Four cases using 4 Fr catheters and data from three of seven operators (due to insufficient case numbers) were excluded. No statistically significant differences in CFR or IMR values were observed among the four operators. Similarly, no significant differences were found between experienced and less-experienced operators (CFR: 1.90 vs. 1.80, p = 0.178; IMR: 19.00 vs. 20.50, p = 0.968). In contrast, CFR values varied significantly by catheter size (5 Fr: 2.20; 6 Fr: 1.80; 7 Fr: 1.50; 8 Fr: 1.70; p < 0.001), whereas IMR remained unaffected. When catheter sizes were dichotomized (small: 5-6 Fr; large: 7-8 Fr), CFR was significantly higher in the small-size group (1.95 vs. 1.65, p = 0.003), while IMR showed no significant difference (20.00 vs. 20.00, p = 0.239).

CONCLUSIONS

While the selection of larger catheter sizes may affect CFR with thermodilution, common sizes and operator techniques show consistent results in practice.

Для цитирования: Hirano Shojiro, Yabe Takayuki, Okubo Ryo, Amano Hideo, Ikeda Takanori. Impact of Procedural Differences on Bolus Thermodilution Method for Assessment of Coronary Microvascular Dysfunction.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2025. DOI: 10.1002/ccd.70085.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.