Optimizing MANTA Closure and Bailout in Large-Bore Femoral Access: Computed Tomography Angiography-Derived Measurements and Rescue Strategies.

Lupu Lior, Hill Andrew P, Haberman Dan, Chitturi Kalyan R, Cellamare Matteo, Zhang Cheng, Sawant Vaishnavi, Abusnina Waiel, Chaturvedi Abhishek, Satler Lowell F, Rogers Toby, Waksman Ron, Ben-Dor Itsik

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

The MANTA (Teleflex, Wayne, PA, USA) facilitates hemostasis in large-bore femoral access. However, precise measurement of the common femoral artery depth is required before arteriotomy upsizing. This measurement may be unavailable in emergencies or bailout scenarios for suture-based closures.

AIMS: This study aimed to evaluate computed tomography (CT) angiography-derived common femoral artery depth measurement as an alternative to the MANTA Depth Locator.

METHODS

This retrospective study at MedStar Washington Hospital Center included all patients with available preprocedural ilio-femoral CT angiography undergoing transfemoral aortic valve reintervention from March 13, 2023, to September 4, 2024. CT measurements of skin-to-artery depth at a designated puncture site, based on the MANTA Instructions for Use 45-degree approach, were compared with intraprocedural MANTA Depth Locator measurements. The acceptable threshold for agreement was defined as ±2 cm to ensure safe in-vessel deployment.

RESULTS

Among 289 patients (mean age 78.2 years, 43.6% female), 91% of CT-derived measurements fell within the 2 cm threshold, demonstrating strong agreement. Logistic regression showed that a higher BMI reduced accuracy, with each unit increase in BMI decreasing the odds of meeting the threshold by 11% (OR 0.89; p < 0.001). For patients with BMI ≤ 30 kg/m2 (n = 189), the agreement increased to 96%.

CONCLUSIONS

CT angiography-derived measurements align well with MANTA Depth Locator measurements, although accuracy declines with increasing BMI. These findings support CT imaging as a reliable alternative to MANTA bailout in large-bore access procedures. A practical algorithm for MANTA bailout techniques is proposed to guide clinical practice.

Для цитирования: Lupu Lior, Hill Andrew P, Haberman Dan, Chitturi Kalyan R, Cellamare Matteo, Zhang Cheng, Sawant Vaishnavi, Abusnina Waiel, Chaturvedi Abhishek, Satler Lowell F, Rogers Toby, Waksman Ron, Ben-Dor Itsik. Optimizing MANTA Closure and Bailout in Large-Bore Femoral Access: Computed Tomography Angiography-Derived Measurements and Rescue Strategies.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2025. DOI: 10.1002/ccd.31661.

Предметные рубрики MeSH