MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/ccd.31396

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2025

Impact of Sheath Type on Vascular and Bleeding Complications After Transcatheter Aortic Valve Replacement: A Post Hoc Analysis From the MARVEL Registry Study.

Pan Emily, Kroon Herbert G, Tonino Pim A L, Amoroso Giovanni, Laine Mika, Christiansen Evald H, Toggweiler Stefan, Ten Berg Jur, Malmberg Markus, Slagboom Ton, Moriyama Noriaki, Terkelsen Christian J, Moccetti Federico, Gheorghe Livia, Bigelow Darra, Webb John, Wood David, Van Mieghem Nicholas, Savontaus Mikko

СтатьяМногоцентровое исследованиеСравнительное исследованиеНаблюдательное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Vascular and bleeding complications remain a concern after transfemoral transcatheter aortic valve replacement (TAVR). The impact of the sheath type on these complications remains unclear.

METHODS

The prospective MARVEL registry study analyzed enrolled 500 patients undergoing large-bore transfemoral procedures and arteriotomy closure with the MANTA vascular closure device from 10 hospitals in Europe and Canada. We stratified these patients according to type of sheath used (expandable or conventional). A propensity-matched analysis was performed using VARC-2 major or minor vascular and bleeding complications as the primary endpoint. The secondary endpoint was time to hemostasis.

RESULTS

We identified 196 propensity-matched pairs. Major vascular complications occurred in 3.6% in the expandable sheath group and 4.1% in the conventional sheath group (p = 1.0). Minor vascular complications occurred in 5.6% in the expandable sheath group and 4.6% in the conventional sheath group (p = 0.819). There were no significant differences in bleeding complications between groups. Time to hemostasis after MANTA closure was significantly shorter in the expandable sheath group (30 vs. 60 s, p < 0.001).

CONCLUSIONS

A propensity-matched analysis demonstrated no significant differences in vascular complication rates with MANTA arteriotomy closure after removal of large bore expandable or conventional sheaths. Time to hemostasis was significantly shorter in the expandable sheath group.

Для цитирования: Pan Emily, Kroon Herbert G, Tonino Pim A L, Amoroso Giovanni, Laine Mika, Christiansen Evald H, Toggweiler Stefan, Ten Berg Jur, Malmberg Markus, Slagboom Ton, Moriyama Noriaki, Terkelsen Christian J, Moccetti Federico, Gheorghe Livia, Bigelow Darra, Webb John, Wood David, Van Mieghem Nicholas, Savontaus Mikko. Impact of Sheath Type on Vascular and Bleeding Complications After Transcatheter Aortic Valve Replacement: A Post Hoc Analysis From the MARVEL Registry Study.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2025. DOI: 10.1002/ccd.31396.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.