Diagnostic reference levels for common paediatric fluoroscopic examinations performed at a dedicated paediatric Australian hospital.

Bibbo Giovanni, Balman Debbie, Linke Rebecca

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Diagnostic reference levels (DRL) of procedures involving ionising radiation are important tools for optimising radiation doses delivered to patients and to identify cases where the levels of dose are unusually high. This is particularly important for paediatric patients undergoing fluoroscopic examinations as these examinations can be associated with a high radiation dose. In this study, a large amount of paediatric fluoroscopic data has been analysed to: establish local DRL, identify the most significant factors determining radiation dose to patients, and modify fluoroscopic techniques to optimise the examination protocols.

METHODS

Paediatric fluoroscopic studies performed at our institution from April 2010 to May 2015 have been retrospectively analysed to determine range, mean, 75th and 95th percentiles of Dose-Area Product (DAP) and fluoroscopic screening time for Micturating Cystourethrography (MCU), Airway, Airway and Swallow, Barium Swallow and Meal, Barium Follow Through and Barium Enema studies.

RESULTS

Currently, no Australian paediatric fluoroscopic DRL data are available for comparison and thus our data can only be compared with international published data. No major changes to examination protocols or modification to fluoroscopic techniques were found necessary as our data compared well with the international published values.

CONCLUSION

The dose delivered to patients depend on a number of factors particularly the experience of the operators. However, DRL are also important, as shown in this study, as they enable best practice by providing feedback to the operators on their performance and benchmarking the institution with other institutions.

Для цитирования: Bibbo Giovanni, Balman Debbie, Linke Rebecca. Diagnostic reference levels for common paediatric fluoroscopic examinations performed at a dedicated paediatric Australian hospital.. Journal of medical imaging and radiation oncology. 2016. DOI: 10.1111/1754-9485.12478.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.