MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/mao.0000000000004699

Медицинская библиотека · Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2025

Perimodiolar Electrode Locations Outperform Lateral Wall Arrays When Controlling for Cochlear Health and Speech Processing Strategy.

Walia Amit, Shew Matthew A, Smetak Miriam, Ortmann Amanda, Durakovic Nedim, Jiramongkolchai Pawina, Lefler Shannon, Herzog Jacques A, Buchman Craig A

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To assess how variations in scala tympani (ST) electrode position affect speech-perception performance, controlling for cochlear health and stimulation strategy.

MATERIALS AND

METHODS

This retrospective cohort study included 98 postlingually deafened adult cochlear implant (CI) recipients at a tertiary referral center. Twenty-one received lateral wall electrodes (CI624; 20 mm) and 77 received perimodiolar electrodes (CI632; 18.4 mm). All implants were positioned with in the ST and programmed with the Advanced Combination Encoder (ACE) strategy. Cochlear health was quantified using pre-insertion round window electrocochleography-total response (ECochG-TR). Postoperative computed tomography with 3-dimensional reconstruction was used to determine electrode modiolar proximity (wrapping factor, WF) and angular insertion depth (AID). The primary outcome was speech perception at 6 months postactivation using CNC word scores.

RESULTS

ECochGTR demonstrated a strong positive correlation with CNC performance (r = 0.61; 95% CI: 0.42 to 0.84). WF correlated weakly and negatively with CNC scores (r = -0.36; 95% CI: -0.53 to -0.16), indicating better outcomes with tighter modiolar proximity. AID showed a weak positive correlation with CNC performance (r = 0.29; 95% CI: 0.08 to 0.47), with deeper insertions associated with improved scores. A regression model using ECochG-TR alone underestimated performance for tightly wrapped and deeply inserted arrays and overestimated performance for lateral wall and shallower insertions. A multivariate model incorporating ECochG-TR, WF, and AID significantly improved predictive accuracy (R² = 0.51; P = 0.001).

CONCLUSIONS

Electrode position within the ST varies considerably, even within the same electrode design; perimodiolar arrays may lie along the lateral wall if over inserted, whereas lateral wall arrays may approach the modiolus depending on cochlear morphology. Incorporating electrode location and cochlear health into predictive models is essential for understanding performance differences between array types. When controlling for ECochG-TR, WF, AID, and programming strategy, perimodiolar electrodes with tighter modiolar proximity outperform lateral wall electrodes at 6 months, emphasizing the importance of these factors in optimizing CI outcomes.

Для цитирования: Walia Amit, Shew Matthew A, Smetak Miriam, Ortmann Amanda, Durakovic Nedim, Jiramongkolchai Pawina, Lefler Shannon, Herzog Jacques A, Buchman Craig A. Perimodiolar Electrode Locations Outperform Lateral Wall Arrays When Controlling for Cochlear Health and Speech Processing Strategy.. Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology. 2025. DOI: 10.1097/mao.0000000000004699.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.