DOI 10.1111/j.1651-2227.2007.00526.x
Correlation between spirometry and impulse oscillometry in children with asthma.
Аннотация
AIM: In certain patients, such as young children or individuals with cerebral palsy or severe mental retardation, it is difficult to perform forced expiratory manoeuvres to measure expiratory flow volume. In such cases, we could evaluate obstructive lung disease through the measurement of airway resistance instead of expiratory flow volume.
METHODS
In this study, we evaluated the correlation of Impulse Oscillometry (IOS) parameters with spirometry values and peak expiratory flow rate (PEFR) measurements to give coherence to IOS recordings in lung function exploration. Total serum IgE levels, total eosinophil counts and specific IgE levels were measured in 48 children with asthma and 66 control subjects, aged 7-15 years of age. IOS, spirometry and PEFR measurements were performed, as well as methacholine challenge. We further analyzed the correlations in atopic asthma, atopic control, nonatopic asthma and nonatopic control groups.
RESULTS
FEV(1) and PEFR showed a significant correlation with impedance and resistance (R) at 5, 10, 20 and 35 Hz, both in atopic asthmatic and in atopic control children. FVC also showed a correlation with impedance and R at 10, 20 and 35 Hz, both in atopic asthmatic and atopic control children. FEF(25-75%) did not show a correlation with resistances.
CONCLUSION
FEV(1), FVC and PEFR were significantly correlated with IOS parameters, in both asthmatic and control subjects, especially for atopic children. IOS could be used as a suitable measure of lung function when spirometry and PEF cannot be performed.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.