MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1186/s13014-026-02893-y

Медицинская библиотека · Radiation oncology (London, England) · 2026

Stereotactic radiosurgery using a self-shielded gyroscopic radiosurgery system for meningiomas: technical aspects and dosimetric results.

Lee Yongsook C, Kotecha Rupesh, Press Robert H, Kalman Noah S, Hall Matthew D, Wieczorek D Jay, Tolakanahalli Ranjini, Mehta Minesh P, McDermott Michael W, Gutierrez Alonso N

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

The ZAP-X® Gyroscopic Radiosurgery® platform (ZAP Surgical Systems, Inc., San Carlos, CA, USA) is a novel self-shielded image-guided radiotherapy device developed for frameless stereotactic radiosurgery (SRS) and fractionated SRS (FSRS) of intracranial and cervical spine lesions. This article presents technical aspects and dosimetric results for SRS and FSRS using the ZAP-X system for meningioma patients.

METHODS

Unique technical aspects of our treatment workflow using the ZAP-X system encompassing computed tomography simulation, treatment planning, independent dose calculations, patient-specific quality assurance (PSQA) and treatment delivery were described. Eighteen single-lesion meningioma patients, who were treated following this workflow, were selected and divided into two groups (Group A [thirteen patients]: 25 Gy in five fractions, FSRS); Group B [five patients]: 15 Gy in a single fraction, SRS). Plan quality metrics, independent dose calculations and PSQA for each group were reviewed and analyzed.

RESULTS

For Group A, mean age (range) was 60 years (35 years - 80 years) and mean planning target volume (PTV) (range) was 4.78 cc (2.10 cc - 9.38 cc). Mean (range) PTV coverage, prescribed isodose line, Paddick conformity index, gradient index, number of isocenters, number of beams, delivery time estimate per fraction and PSQA gamma passing rate were 99.5% (99.5% - 99.6%), 53.5% (50.1% - 61.3%), 0.89 (0.83-0.93), 2.64 (2.54-2.88), 11.8 (5-19), 208.5 (91-332), 43.6 min (24.5 -61.0 min) and 95.0% (91.5% - 98.4%), respectively. For Group B, corresponding values were 70 years (62 years - 75 years), 1.12 cc (0.77 cc - 1.99 cc), 99.5% (99.5% - 99.5%), 56.4% (50.6% - 62.8%), 0.89 (0.85-0.93), 2.95 (2.63-3.22), 7.2 (5-9), 102.4 (81-132), 33.9 min (27.3 -38.9 min) and 95.3% (91.1% - 100.0%). For all beams in both groups, independent dose calculations were < ± 0.02% of treatment planning system dose calculations.

CONCLUSIONS

Our experience with SRS and FSRS using the ZAP-X system for meningioma patients demonstrated that a typical external beam radiotherapy workflow was applicable with unique technical aspects, and treatment plans met our institutional planning goals and were concordant with independent dose calculations and PSQA.

Для цитирования: Lee Yongsook C, Kotecha Rupesh, Press Robert H, Kalman Noah S, Hall Matthew D, Wieczorek D Jay, Tolakanahalli Ranjini, Mehta Minesh P, McDermott Michael W, Gutierrez Alonso N. Stereotactic radiosurgery using a self-shielded gyroscopic radiosurgery system for meningiomas: technical aspects and dosimetric results.. Radiation oncology (London, England). 2026. DOI: 10.1186/s13014-026-02893-y.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.