The Annals of thoracic surgery · 1995

Initial experience with the Cosgrove-Edwards Annuloplasty System.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Cosgrove D M, Arcidi J M, Rodriguez L, Stewart W J, Powell K, Thomas J D

Аннотация

BACKGROUND: The mitral valve has a nonplanar shape and a sphincter action. Pathologic dilatation occurs along the posterior annulus. To preserve the physiologic function and correct annular dilatation, we developed an annuloplasty system that is universally flexible and produces a measured plication of the posterior annulus (Cosgrove-Edwards Annuloplasty System).

METHODS: The results of 150 consecutive mitral valve repairs using this system were analyzed. Mean age was 58 +/- 13 years; 59% were men. The cause of the valve disease was degenerative in 75% of the patients, rheumatic in 13%, ischemic in 8%, and infectious in 4%. Associated procedures were performed in 61 patients (41%).

RESULTS: Echocardiographic mitral regurgitation decreased from 3.7 +/- 0.6 before repair to 0.2 +/- 0.4 after repair (p < 0.0001). There were no hospital deaths and no cases of hemodynamically significant systolic anterior motion or other annuloplasty-related complications. Follow-up was 100% complete at a mean of 3.1 +/- 3.6 months. There were three late deaths, three transient ischemic attacks, and one episode of endocarditis. Five patients (3.3%) have undergone reoperation for recurrent mitral insufficiency; no reoperations were related to the annuloplasty system. At a mean of 9 months, three-dimensional reconstruction of the mitral annulus from multiple echocardiographic images confirmed the nonplanar shape and sphincter mechanism of the annulus. Annular orifice area decreased 19% during the cardiac cycle from a mean of 10.3 cm2 in diastole to 8.6 cm2 in systole.

CONCLUSIONS: This annuloplasty system is effective for repair of insufficiency secondary to all causes, preserves physiologic annulus function, and is associated with a low incidence of valve-related complication.