DOI 10.1097/mao.0000000000001820
Benefits of Bimodal Hearing With Cochlear and Middle Ear Implants: Preliminary Results in Four Patients.
Аннотация
OBJECTIVE
Conventional hearing aids have some limitations in overcoming a large air-bone gap or in cases of severe high-frequency hearing loss. The authors aimed to evaluate the benefit of a new bimodal hearing configuration combining cochlear implantation (CI) and middle ear implant (MEI) in patients with severe mixed conductive and ski-slope hearing loss.
STUDY
DESIGN
Retrospective case review.
SETTING
Tertiary referral center.
PATIENTS: Four patients with severe to profound hearing loss, who underwent CI in one ear and MEI in the other, were enrolled.
INTERVENTION: Audiological outcomes were assessed at least 6 months after the MEI/CI operation. Pure-tone audiometry (PTA) in unaided, CI-aided, and MEI-aided conditions were measured.
MAIN OUTCOME MEASURES: Average threshold changes in bands of frequencies (<1 kHz, ≥1 kHz) were compared between MEI-aided and HA-aided conditions. The Korean version of the Hearing-in-Noise Test (K-HINT), and speech perception score in noisy and quiet conditions were evaluated in the bimodal configuration (i.e., MEI with CI).
RESULTS
MEI-aided PTA was especially increased in high-frequency areas (≥1 kHz). Speech perception in noisy and quiet conditions demonstrated better scores in the bimodal configuration. K-HINT also demonstrated better scores in the bimodal configuration.
CONCLUSIONS
There is an increasing number of patients with unilateral CI and residual hearing in the contralateral ear. The benefits of a new bimodal hearing configuration with CI and MEI were demonstrated in patients with severe high-frequency hearing loss or mixed conductive hearing loss in the contralateral ear.