Accuracy and Interpretation of Transcutaneous Carbon Dioxide Monitoring in Critically Ill Children.

Setar Leah, Lee Jessica G, Sanchez-Pinto L Nelson, Coates Bria M

СтатьяНаблюдательное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

Transcutaneous carbon dioxide (Tc co2 ) monitoring can noninvasively assess ventilation by estimating carbon dioxide ( CO2 ) levels in the blood. We aimed to evaluate the accuracy of Tc co2 monitoring in critically ill children by comparing it to the partial pressure of arterial carbon dioxide (Pa co2 ). In addition, we sought to determine the variation between Tc co2 and Pa co2 acceptable to clinicians to modify patient care and to determine which patient-level factors may affect the accuracy of Tc co2 measurements.

DESIGN

Retrospective observational cohort study.

SETTING

Single, quaternary care PICU from July 1, 2012, to August 1, 2020.

PATIENTS: Included participants were admitted to the PICU and received noninvasive ventilation support (i.e., continuous or bilevel positive airway pressure), conventional mechanical ventilation, or high-frequency oscillatory or percussive ventilation with Tc co2 measurements obtained within 15 minutes of Pa co2 measurement.

INTERVENTIONS: None.

MEASUREMENTS AND MAIN

RESULTS

Three thousand four hundred seven paired arterial blood gas and Tc co2 measurements were obtained from 264 patients. Bland-Altman analysis revealed a bias of -4.4 mm Hg (95% CI, -27 to 18.3 mm Hg) for Tc co2 levels against Pa co2 levels on the first measurement pair for each patient, which fell within the acceptable range of ±5 mm Hg stated by surveyed clinicians, albeit with wide limits of agreement. The sensitivity and specificity of Tc co2 to diagnose hypercarbia were 93% and 71%, respectively. Vasoactive-Infusion Score (VIS), age, and self-identified Black/African American race confounded the relationship between Tc co2 with Pa co2 but percent fluid overload, weight-for-age, probe location, and severity of illness were not significantly associated with Tc co2 accuracy.

CONCLUSIONS

Tc co2 monitoring may be a useful adjunct to monitor ventilation in children with respiratory failure, but providers must be aware of the limitations to its accuracy.

Для цитирования: Setar Leah, Lee Jessica G, Sanchez-Pinto L Nelson, Coates Bria M. Accuracy and Interpretation of Transcutaneous Carbon Dioxide Monitoring in Critically Ill Children.. Pediatric critical care medicine : a journal of the Society of Critical Care Medicine and the World Federation of Pediatric Intensive and Critical Care Societies. 2024. DOI: 10.1097/pcc.0000000000003564.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.