DOI 10.1097/mao.0b013e31816fdc90

Медицинская библиотека · Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2008

Percutaneous versus transcutaneous bone conduction implant system: a feasibility study on a cadaver head.

Håkansson Bo, Eeg-Olofsson Måns, Reinfeldt Sabine, Stenfelt Stefan, Granström Gösta

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

Percutaneous bone-anchored hearing aid (BAHA) is an important rehabilitation alternative for patients who have conductive or mixed hearing loss. However, these devices use a percutaneous and bone-anchored implant that has some drawbacks reported. A transcutaneous bone conduction implant system (BCI) is proposed as an alternative to the percutaneous system because it leaves the skin intact. The BCI transmits the signal to a permanently implanted transducer with an induction loop system through the intact skin. The aim of this study was to compare the electroacoustic performance of the BAHA Classic-300 with a full-scale BCI on a cadaver head in a sound field. The BCI comprised the audio processor of the vibrant sound bridge connected to a balanced vibration transducer (balanced electromagnetic separation transducer).

METHODS

Implants with snap abutments were placed in the parietal bone (Classic-300) and 15-mm deep in the temporal bone (BCI). The vibration responses at the ipsilateral and contralateral cochlear promontories were measured with a laser Doppler vibrometer, with the beam aimed through the ear canal.

RESULTS

Results show that the BCI produces approximately 5 dB higher maximum output level and has a slightly lower distortion than the Classic-300 at the ipsilateral promontorium at speech frequencies. At the contralateral promontorium, the maximum output level was considerably lower for the BCI than for the Classic-300 except in the 1-2 kHz range, where it was similar.

CONCLUSION

Present results support the proposal that a BCI system can be a realistic alternative to a BAHA.

Для цитирования: Håkansson Bo, Eeg-Olofsson Måns, Reinfeldt Sabine, Stenfelt Stefan, Granström Gösta. Percutaneous versus transcutaneous bone conduction implant system: a feasibility study on a cadaver head.. Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology. 2008. DOI: 10.1097/mao.0b013e31816fdc90.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.