MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3109/00016489.2013.808764

Медицинская библиотека · Acta oto-laryngologica · 2013

Use of bone anchoring device in electromagnetic computer-assisted navigation in lateral skull base surgery.

Bernardeschi Daniele, Nguyen Yann, Villepelet Aude, Ferrary Evelyne, Mazalaigue Stephane, Kalamarides Michel, Sterkers Olivier

СтатьяПеревод пока не подготовлен

Аннотация

CONCLUSION

The use of the bone anchoring device associated with a fiducial marker, both fixed close to the operating field, improves the reproducibility and effectiveness of the computer-assisted navigation in lateral skull base surgery.

OBJECTIVES

Computer-assisted navigation in lateral skull base surgery using the electromagnetic system Digipointeur(®) needs an external fiducial marker (titanium screw) close to the operating field to increase position accuracy (PA) to about 1 mm. Displacement of the emitter placed in the mouth (Buccostat(®)) induces a drift of the system, leading to at least 20% of unsuccessful procedures. The aim of this study was to evaluate the PA, stability, and reproducibility of computer-assisted navigation in lateral skull base surgery using a bone anchoring device to provide a fixed registration system near the operating field.

METHODS

Forty patients undergoing a lateral skull base procedure with the Digipointeur(®) system performed with both the titanium screw and bone anchoring device were included in this prospective study. They were divided in two groups. In the first one (n = 9), the PA was measured before and after screw registration for five intratemporal landmarks, during a translabyrinthine approach. In the second group (n = 31), all lateral skull base procedures were included and the PA was evaluated visually by the surgeon on different landmarks of the approaches as well as the stability of the system.

RESULTS

In the first group, the PA was 7.08 ± 0.59 mm and 0.77 ± 0.17 mm (mean ± SEM, p < 0.0001) before and after screw registration, respectively. In the second group, the PA was considered as accurate by the surgeon in all cases and no drift of the system was observed. Computer-assisted surgery was never abandoned due to increased stability of the bone-anchored emitter.

Для цитирования: Bernardeschi Daniele, Nguyen Yann, Villepelet Aude, Ferrary Evelyne, Mazalaigue Stephane, Kalamarides Michel, Sterkers Olivier. Use of bone anchoring device in electromagnetic computer-assisted navigation in lateral skull base surgery.. Acta oto-laryngologica. 2013. DOI: 10.3109/00016489.2013.808764.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.