MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.ijporl.2011.11.016

Медицинская библиотека · International journal of pediatric otorhinolaryngology · 2012

Auditory brainstem implant outcomes and MAP parameters: report of experiences in adults and children.

Goffi-Gomez Maria Valéria Schmidt, Magalhães Ana Tereza, Brito Neto Rubens, Tsuji Robinson Koji, Gomes Marcos de Queiroz Telles, Bento Ricardo Ferreira

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

UNLABELLED: The auditory brainstem implant (ABI) was first developed to help neurofibromatosis type 2 patients. Recently, its use has been recently extended to adults with non-tumor etiologies and children with profound hearing loss who were not candidates for a cochlear implant (CI). Although the results has been extensively reported, the stimulation parameters involved behind the outcomes have received less attention.

OBJECTIVE

The aim of this study is to describe the audiologic outcomes and the MAP parameters in ABI adults and children at our center.

METHODS

Retrospective chart review. Five adults and four children were implanted with the ABI24M from September 2005 to June 2009. In the adult patients, four had Neurofibromatosis type 2, and one had postmeningitic deafness with complete ossification of both cochleae. Three of the children had cochlear malformation or dysplasia, and one had complete ossified cochlea due to meningitis. Map parameters as well as the intraoperative electrical auditory brainstem responses were collected. Evaluation was performed with at least six months of device use and included free-field hearing thresholds, speech perception tests in the adult patients and for the children, the Infant-Toddler Meaningful Auditory Integration Scale (IT-MAIS) and (ESP) were used to evaluate the development of auditory skills, besides the MUSS to evaluate.

RESULTS

The number of active electrodes that did not cause any non-auditory sensation varied from three to nineteen. All of them were programmed with SPEAK strategy, and the pulse widths varied from 100 to 300 μs. Free-field thresholds with warble tones varied from very soft auditory sensation of 70 dBHL at 250 Hz to a pure tone average of 45 dBHL. Speech perception varied from none to 60% open-set recognition of sentences in silence in the adult population and from no auditory sensation at all to a slight improvement in the IT-MAIS/MAIS scores.

CONCLUSION

We observed that ABI may be a good option for offering some hearing attention to both adults and children. In children, the results might not be enough to ensure oral language development. Programming the speech processor in children demands higher care to the audiologist.

Для цитирования: Goffi-Gomez Maria Valéria Schmidt, Magalhães Ana Tereza, Brito Neto Rubens, Tsuji Robinson Koji, Gomes Marcos de Queiroz Telles, Bento Ricardo Ferreira. Auditory brainstem implant outcomes and MAP parameters: report of experiences in adults and children.. International journal of pediatric otorhinolaryngology. 2012. DOI: 10.1016/j.ijporl.2011.11.016.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.