DOI 10.1097/00003246-199711000-00015

Медицинская библиотека · Critical care medicine · 1997

Lithium dilution cardiac output measurement: a comparison with thermodilution.

Linton R, Band D, O'Brien T, Jonas M, Leach R

Клиническое исследованиеСравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

To compare the results of cardiac output measurements obtained by lithium dilution and thermodilution.

DESIGN

Case series, observational study.

SETTING

High-dependency postoperative unit and intensive care unit of a teaching hospital.

PATIENTS: Forty patients were studied. Thirty-four patients had undergoing heart surgery requiring cardiopulmonary bypass within the previous 2 days; the diagnoses in the other patients were myocardial infarct (n = 2), septicemia (n = 2), adult respiratory distress syndrome, and pericardectomy.

INTERVENTIONS: Cardiac output was measured five times in each patient, using lithium dilution (single measurement) and bolus thermodilution (series of three to six measurements according to standard clinical practice, taking the average of the closest three). In a subgroup of 14 patients, cardiac output was also measured using "continuous thermodilution."

MEASUREMENTS AND MAIN

RESULTS

Comparing lithium dilution with bolus thermodilution, the mean of the differences (lithium dilution-thermodilution) was -0.25 +/- 0.46 [SD] L/min. Linear regression analysis gave y = 0.31 + 0.89x (r2 = .94) for lithium dilution vs. thermodilution.

CONCLUSIONS

The overall agreement between the two methods was good. The variability of the thermodilution measurements was greater than that of the lithium dilution measurements. The lithium dilution method is at least as accurate as bolus thermodilution and, since pulmonary artery catheterization is not needed, it has the advantages of being safe and quick to perform.

Для цитирования: Linton R, Band D, O'Brien T, Jonas M, Leach R. Lithium dilution cardiac output measurement: a comparison with thermodilution.. Critical care medicine. 1997. DOI: 10.1097/00003246-199711000-00015.