Accurate delivery of low flow oxygen is essential for the management of infants with chronic respiratory failure. Twenty-seven examples of one low flow oxygen meter were tested. Measured flows were found to differ greatly from indicated settings, particularly at 250 mL/min settings. Knowledge of such errors as well as improved quality control are essential for appropriate management of oxygen therapy.