DOI 10.1177/03000605261418843

Медицинская библиотека · The Journal of international medical research · 2026

Evaluation of repeatability and reproducibility of central and peripheral axial length measurements in children using the novel Swan optical biometer.

Wang Qingxia, Yu Huan, Zhou Dan, Ma Xiaoyun

СтатьяПеревод пока не подготовлен

Аннотация

ObjectiveTo evaluate the repeatability and reproducibility of central axial length and peripheral axial length measurements in children using the Swan 660 M.MethodsThis cross-sectional study was conducted from August to September 2024. A total of 60 children (6-15 years; no significant ocular diseases) were recruited from Zhoupu Hospital, Shanghai (30 males and 30 females; mean age 10.67 ± 2.47 years), and only right eyes were included in the analysis. The Swan 660 M was used to measure central axial length and peripheral axial length. Measurement variability across operator and time was evaluated using the intra-group SD, test-retest repeatability, coefficient of variation, and intraclass correlation coefficient.ResultsThe Swan 660 M demonstrated good repeatability and reproducibility for the measurements of central axial length and most peripheral axial length points, including Nasal 1, Temporal 1, Temporal 2, and Temporal 3 (coefficient of variation <5%, intraclass correlation coefficient >0.9). However, measurement reliability was lower for certain peripheral axial length points, such as Nasal 5 (intraclass correlation coefficient <0.75). The device also showed high stability in the measurements of central axial length (intraclass correlation coefficient = 0.999).ConclusionsThe Swan 660 M demonstrates high repeatability and reproducibility for central axial length and most peripheral axial length measurements in children, suggesting its potential application in tracking the progression of myopia and investigating the relationships between peripheral retinal morphology and refractive status in this population.

Для цитирования: Wang Qingxia, Yu Huan, Zhou Dan, Ma Xiaoyun. Evaluation of repeatability and reproducibility of central and peripheral axial length measurements in children using the novel Swan optical biometer.. The Journal of international medical research. 2026. DOI: 10.1177/03000605261418843.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.