Аннотация

OBJECTIVE

To describe the first cochlear array insertions using a robot-assisted technique, with different types of straight or precurved electrode arrays, compared with arrays manually inserted into the cochlea.

STUDY

DESIGN

Retrospective review.

SETTING

Tertiary otologic center.

PATIENTS: Twenty cochlear implantations in the robot-assisted group and 40 in the manually inserted group.

INTERVENTIONS: Cochlear implantations using a robot-assisted technique (RobOtol) with straight (eight Cochlear CI522/622, and eight Advanced Bionics Hifocus Slim J) or precurved (four Advanced Bionics Hifocus Mid-Scala) matched to manual cochlear implantations. Three-dimensional reconstruction images of the basilar membrane and the electrode array were obtained from pre- and postimplantation computed tomography.

MAIN OUTCOME MEASURES: Rate and localization of scalar translocations.

RESULTS

For straight electrode arrays, scalar translocations occurred in 19% (3/16) of the robot-assisted group and 31% (10/32) of the manually inserted group. Considering the number of translocated electrodes, this was lower in the robot-assisted group (7%) than in the manually inserted group (16%) (p < 0.0001, χ2 test). For precurved electrode arrays, scalar translocations occurred in 50% (2/4) of the robot-assisted group and 38% (3/8) of the manually inserted group.

CONCLUSION

This study showed a safe and reliable insertion of different electrode array types with a robot-assisted technique, with a less traumatic robotic insertion of straight electrode arrays when compared with manual insertion.

Для цитирования: Daoudi Hannah, Lahlou Ghizlene, Torres Renato, Sterkers Olivier, Lefeuvre Vincent, Ferrary Evelyne, Mosnier Isabelle, Nguyen Yann. Robot-assisted Cochlear Implant Electrode Array Insertion in Adults: A Comparative Study With Manual Insertion.. Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology. 2021. DOI: 10.1097/mao.0000000000003002.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.