Is There Any Correlation between Spread of Excitation Width and the Refractory Properties of the Auditory Nerve in Cochlear Implant Users?
Аннотация
BACKGROUND
The spread of excitation (SOE) and auditory nerve recovery function (REC) are objective measures recorded by neural response telemetry and may interfere in cochlear implant (CI) stimulation.
OBJECTIVE
To analyze and correlate SOE with the refractory periods in subjects with pre- and postlingual deafness implanted with different electrode arrays.
METHODS
This was a retrospective study of 323 ears separated by perimodiolar or straight arrays and by pre- or postlingually deaf recipients. Measures were collected intraoperatively on electrode 11. The SOE width was measured in millimeters at the 0.75 point of the curve, and the relative (tau) and absolute (t0) refractory periods were measured in microseconds.
RESULTS
There was a statistical correlation between the SOE and the t0 in the patients with postlingual deafness implanted with the perimodiolar array. The SOE width was statistically different between the straight and perimodiolar arrays and between the pre- and postlingual groups in the perimodiolar array. Tau was statistically different between the pre- and postlingual groups with the straight array and the t0, between the pre- and postlingual groups with the perimodiolar array. Neural response threshold and amplitude of the neural response were not statistically different among groups.
CONCLUSION
There was a correlation between SOE width and t0 only in patients with acquired deafness. The findings suggest that different factors influence SOE and REC, considering SOE is different according to the electrode array and REC being different according the onset of deafness.