Fluid Dynamics of a Novel Micro-Fistula Implant for the Surgical Treatment of Glaucoma.

Sheybani Arsham, Reitsamer Herbert, Ahmed Iqbal Ike K

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: The purpose of this study was to describe the fluidics of a novel non-valved glaucoma implant designed to prevent hypotony and compare the fluidics of this device with two commonly used non-valved glaucoma devices.

METHODS

The XEN 45 micro-fistula implant was designed to limit hypotony by virtue of its length and width according to the Hagen-Poiseuille equation. Flow testing was performed using a syringe pump and pressure transducer at multiple flow rates. The pressure differentials across the XEN implant, the Ex-Press implant, and 10 mm of silicone tubing from a Baerveldt implant at a physiologic flow rate (2.5 μL/min) were extrapolated.

RESULTS

The XEN 45 achieved a steady-state pressure calculated at 7.56 mm Hg at 2.5 μL/min. At the same flow rate, the Ex-Press device and Baerveldt tubing reached steady-state pressures of 0.09 and 0.01 mm Hg, respectively.

CONCLUSIONS

Under flow testing, the XEN micro-fistula implant was able to maintain backpressure above numerical hypotony levels without the use of complex valve systems. This is due to the XEN implant's design, derived from the principles that dictate Newtonian fluids.

Для цитирования: Sheybani Arsham, Reitsamer Herbert, Ahmed Iqbal Ike K. Fluid Dynamics of a Novel Micro-Fistula Implant for the Surgical Treatment of Glaucoma.. Investigative ophthalmology & visual science. 2015. DOI: 10.1167/iovs.15-16625.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.