Ehlayel Abdulla M, Okunowo Oluwatimilehin, Dutt Mohini, Howarth Kathryn, Zemel Babette S, Poznick Laura, Morgan Xenia, Denburg Michelle R, Copelovitch Lawrence, Back Susan J, Otero Hansel J, Hartung Erum A
Аннотация
BACKGROUND: Fluid overload is associated with morbidity and mortality in children receiving dialysis. Accurate clinical assessment is difficult, and using deuterium oxide (D2O) to measure total body water (TBW) is impractical. We investigated the use of ultrasound (US), bioimpedance spectroscopy (BIS), and anthropometry to assess fluid removal in children receiving maintenance hemodialysis (HD).
METHODS: Participants completed US, BIS, and anthropometry immediately before and 1-2 h after HD for up to five sessions. US measured inferior vena cava (IVC) diameter, lung B-lines, muscle elastography, and dermal thickness. BIS measured the volume of extracellular (ECF) and intracellular (ICF) fluid. Anthropometry included mid-upper arm, calf and ankle circumferences, and triceps skinfold thickness. D2O was performed once pre-HD. We assessed the change in study measures pre- versus post-HD, and the correlation of change in study measures with percent change in body weight (%∆BW). We also assessed the agreement between TBW measured by BIS and D2O.
RESULTS: Eight participants aged 3.4-18.5 years were enrolled. Comparison of pre- and post-HD measures showed significant decrease in IVC diameters, lung B-lines, dermal thickness, BIS %ECF, mid-upper arm circumference, ankle, and calf circumference. Repeated measures correlation showed significant relationships between %∆BW and changes in BIS ECF (rrm =0.51, 95% CI 0.04, 0.80) and calf circumference (rrm=0.80, 95% CI 0.51, 0.92). BIS TBW correlated with D2O TBW but overestimated TBW by 2.2 L (95% LOA, -4.75 to 0.42).
CONCLUSION: BIS and calf circumference may be helpful to assess changes in fluid status in children receiving maintenance HD. IVC diameter, lung B-lines and dermal thickness are potential candidates for future studies.
METHODS: Participants completed US, BIS, and anthropometry immediately before and 1-2 h after HD for up to five sessions. US measured inferior vena cava (IVC) diameter, lung B-lines, muscle elastography, and dermal thickness. BIS measured the volume of extracellular (ECF) and intracellular (ICF) fluid. Anthropometry included mid-upper arm, calf and ankle circumferences, and triceps skinfold thickness. D2O was performed once pre-HD. We assessed the change in study measures pre- versus post-HD, and the correlation of change in study measures with percent change in body weight (%∆BW). We also assessed the agreement between TBW measured by BIS and D2O.
RESULTS: Eight participants aged 3.4-18.5 years were enrolled. Comparison of pre- and post-HD measures showed significant decrease in IVC diameters, lung B-lines, dermal thickness, BIS %ECF, mid-upper arm circumference, ankle, and calf circumference. Repeated measures correlation showed significant relationships between %∆BW and changes in BIS ECF (rrm =0.51, 95% CI 0.04, 0.80) and calf circumference (rrm=0.80, 95% CI 0.51, 0.92). BIS TBW correlated with D2O TBW but overestimated TBW by 2.2 L (95% LOA, -4.75 to 0.42).
CONCLUSION: BIS and calf circumference may be helpful to assess changes in fluid status in children receiving maintenance HD. IVC diameter, lung B-lines and dermal thickness are potential candidates for future studies.