MedElevator Войти в исследования
← Поиск публикаций

DOI 10.5664/jcsm.10726

Медицинская библиотека · Journal of clinical sleep medicine : JCSM : official publication of the American Academy of Sleep Medicine · 2023

Home monitoring for clinically suspected obstructive sleep apnea in pregnancy.

Wang Jingyu, Zhang Chi, Xu Liyue, Wei Jun, Yang Jingjing, Strohl Anna E, Yi Huijie, Liu Xinran, Zhang Linyan, Zhao Rui, Hong Shenda, Zhou Bing, Zhao Long, Zhang Xueli, Dong Xiaosong, Strohl Kingman P, Wang Jianliu, Liu Guoli, Han Fang

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

STUDY

OBJECTIVES

To determine if a home sleep apnea test (HSAT) using a type III portable monitor (PM), Nox-T3 (Nox Medical, Inc., Reykjavik, Iceland), detects obstructive sleep apnea in pregnant women.

METHODS

Ninety-two pregnant women (34.5 ± 4.3 years; gestational age 25.4 ± 8.9 weeks; body mass index 29.9 ± 4.7 kg/m2) with suspected obstructive sleep apnea underwent HSAT with the Nox-T3 PM followed by overnight polysomnography (PSG) and PM recording simultaneously in the laboratory within 1 week. PMs were scored automatically and manually using a 3% criteria and compared with PSGs scored by following guidelines.

RESULTS

Apnea-hypopnea indexes were 8.56 ± 10.42, 8.19 ± 13.79, and 8.71 ± 14.19 events/h on HSAT, in-laboratory PM recording, and PSG (P = .955), respectively. Bland-Altman analysis of the apnea-hypopnea index on PSG vs HSAT showed a mean difference (95% confidence interval) of -0.15 (-1.83, 1.53); limits of agreement (± 2 SD) were -16.26 to 16.56 events/h. Based on a threshold apnea-hypopnea index ≥ 5 events/h, HSAT had 91% sensitivity, 85% specificity, 84% positive-predictive value, and 92% negative-predictive value compared with PSG. When comparing the simultaneous recordings, closer agreement was observed. Automated vs manual analysis of PM showed no significant difference.

CONCLUSIONS

A type III PM had an acceptable failure rate and high diagnostic performance operating as a reasonable alternative for in-laboratory PSG in pregnant women.

CITATION: Wang J, Zhang C, Xu L, et al. Home monitoring for clinically suspected obstructive sleep apnea in pregnancy. J Clin Sleep Med. 2023;19(11):1951-1960.

Для цитирования: Wang Jingyu, Zhang Chi, Xu Liyue, Wei Jun, Yang Jingjing, Strohl Anna E, Yi Huijie, Liu Xinran, Zhang Linyan, Zhao Rui, Hong Shenda, Zhou Bing, Zhao Long, Zhang Xueli, Dong Xiaosong, Strohl Kingman P, Wang Jianliu, Liu Guoli, Han Fang. Home monitoring for clinically suspected obstructive sleep apnea in pregnancy.. Journal of clinical sleep medicine : JCSM : official publication of the American Academy of Sleep Medicine. 2023. DOI: 10.5664/jcsm.10726.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.