DOI 10.1080/08164622.2021.1878854
Evaluation of tear meniscus height using different clinical methods.
Аннотация
Clinical relevance: The height of the tear meniscus (TMH) is a generally accepted method to evaluate tear film volume, especially in dry eye diagnoses and management.
Background
The purpose of this study was to evaluate the ability of different methods to measure tear meniscus height accurately and repeatably.
Methods
Lower TMH of 20 volunteers (26.8 ± 5.6 years) was measured by two observers (OI and OII) using optical coherence tomography (OCT), slitlamp microscope image analysis, and with a reticule at low (8x) and high (32x) magnification. TMH was also evaluated by both observers by comparing TMH to thickness of the lid margin (lid-ratio; grade 0: TMH 1/2 lid margin thickness; grade 1: 1/3; grade 2: 1/4; grade 3: 1/5; grade 4: 1/6) and to the number of eyelashes fitting in the tear meniscus. Differences between observers were analysed by paired-t-test. Differences between OCT-TMH and other methods were analysed by ANOVA, and inter-observer repeatability by intra-class-correlation-coefficient (ICC). The ability to predict OCT-TMH was calculated by receiver operative characteristic (ROC) curve analysis.
Results
There was no significant difference between OI and OII in all methods except of the eyelash-count-method (p = 0.008). For OI, TMH measured using a reticule at 8x (0.20 ± 0.05 mm) was significantly lower than OCT-TMH (0.24 ± 0.07 mm) (p = 0.032) but not at 32x (0.22 ± 0.01 mm; p = 0.435). TMH evaluated by the image software of the slitlamp (0.20 ± 0.05 mm) was significantly lower than OCT-TMH (p = 0.022). The lid-ratio-method and eyelash-count-method resulted in grades of 2.35 ± 1.22 and 2.85 ± 0.81, respectively. ROC analyses showed that only the 8x and the 32x magnification method could discriminate between normal and abnormal OCT-TMH. OCT had the best repeatability (ICC = 0.88; p < 0.001) followed by reticule using 32x magnification (ICC = 0.70; p = 0.004).
Conclusion
The most reliable method to measure TMH was OCT followed by slitlamp using a reticule. TMH cannot be reliably evaluated by comparing it against lid margin thickness or number of eyelashes.