Reduced-Dose Chest CTA for the Detection of Pulmonary Arteriovenous Malformations in Pediatric Patients With Hereditary Hemorrhagic Telangiectasia.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Saadeh Zaid, Demirel Nadir, Horst Kelly K, Iyer Vivek N, Koo Chi Wan, Larson Nicholas B, McCollough Cynthia H, Oo Daniel, Tandon Yasmeen K, Thorne Jamison E, Zhou Zhongxing, Yu Lifeng, Hull Nate C

Аннотация

OBJECTIVES: To determine the feasibility of reduced-dose chest computed tomographic angiography (CTA) with convolutional neural network (CNN) denoising for detecting pulmonary arteriovenous malformations (pAVMs) in children with hereditary hemorrhagic telangiectasia (cwHHT).

METHODS: Fifteen cwHHT underwent a chest CTA (ie, a controlled "study" dose). Noise was inserted to simulate a quarter dose (QD) exam. Images were reconstructed using iterative reconstruction (IR) and our self-trained CNN denoising model. For each case, 3 sets of images were created: study dose (SD)+IR, QD+IR, and QD+CNN. Two thoracic radiologists independently scored each set to assess quality, spatial resolution, artifacts, and the presence of pAVMs using 4-level ordinal scales. Quantitative assessments of image quality were performed using contrast-to-noise ratios (CNRs) with comparisons made between the experimental conditions.

RESULTS: Thirteen of the 15 patients recruited with hereditary hemorrhagic telangiectasia (mean age: 9.3±4.5 y) were positive for pAVM by transthoracic contrast echocardiography. The sensitivities using QD+CNN were 0.85 and 1.00 for readers 1 and 2, respectively. This was compared with 0.69 and 0.84 using QD+IR versus 0.85 and 0.92 for SD+IR. Inter-reader agreement for pAVM detection utilizing QD+CNN was moderate and resulted in kappa=0.59 ( P =0.012). The subjective assessments for QD+CNN were comparable to the SD technique. Regression analysis of reader scores revealed improved quality in QD+CNN versus QD+IR ( P =0.001). Similarly, the QD+CNN condition demonstrated the highest CNRs.

CONCLUSIONS: Reduced-dose chest CTA with CNN denoising provides a level of sensitivity comparable to standard dose CTA and high CNRs for the detection of pAVMs in cwHHT.