DOI 10.1016/j.athoracsur.2020.12.051
Melody Valve Fracture Causing Mitral Stenosis: Novel Solution for Transapical Valve-in-Valve.
Аннотация
The use of Melody valves in the mitral position has been introduced in clinical practice. Stent fracture is a recognized complication of Melody valve implantation in the pulmonary position; however, reports in the mitral position are rare. We present the case of an 8-year-old boy in whom complete fracture of the proximal stent struts occurred, causing acute severe mitral stenosis, and in whom urgent hybrid transapical Melody valve implantation in the fractured Melody valve was performed successfully using a novel modified technique. This modification simplified the implantation, led to reduced time of the procedure, and minimized hemodynamic instability.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.