m+Medelevator.RU
← Поиск публикаций
Journal of cataract and refractive surgery · 2003

Optical and ultrasound measurement of axial length and anterior chamber depth for intraocular lens power calculation.

Клиническое исследованиеКонтролируемое клиническое исследованиеСтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Németh János, Fekete Orsolya, Pesztenlehrer Norbert

Аннотация

PURPOSE: To evaluate the precision, reproducibility, and applicability of an optical method based on partial coherence interferometry for intraocular lens (IOL) power calculation.

SETTING: Ultrasound laboratory of a university eye hospital.

METHODS: A prospective comparison of measurements made by the IOLMaster optical instrument (Carl Zeiss) and Ultrascan Digital 2000 contact ultrasound A-scan (Alcon) for IOL calculations was performed. Examined were 255 eyes of 134 persons (204 phakic, 47 pseudophakic, and 4 aphakic). The mean age of the patients was 67.9 years (range 7 to 94 years).

RESULTS: The IOLMaster measurements were successful in more than 80% of cases: in 82%, 99%, and 99% for axial length (AL), anterior chamber depth (ACD), and keratometry measurements, respectively. The reproducibility of the AL and ACD measurements was very high (coefficient of variation 0.13% and 2.20%, respectively). The AL and ACD values were significantly larger with the IOLMaster (P <.001) than with the Ultrascan Digital 2000. The correlation between ultrasound and optical AL measurements was high (r = 0.985; P <.001); however, there was no correlation between ACD measurements (r = 0.079; P =.397). The corneal refractive power measurements of a Javal-type keratometer and the IOLMaster were highly correlated (r = 0.955; P <.001), with a mean difference of 0.2 diopter.

CONCLUSIONS: The results show that measurements for IOL calculation are easy and precise with the optical method. It is a noncontact method, so no anesthesia is needed and there is no risk of infection.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты