Long-term outcomes of endovascular aneurysm repair according to instructions for use adherence status.

Hahl Tilda, Protto Sara, Järvenpää Valtteri, Uurto Ilkka, Väärämäki Suvi, Suominen Velipekka

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

Endovascular aneurysm repair (EVAR) has become the standard treatment for abdominal aortic aneurysms (AAAs). Endovascular device manufacturers have defined specific anatomic criteria for the aneurysm characteristics that should be observed as instructions for use (IFU) for specific grafts. In clinical practice, the prevalence of performing EVAR outside the IFU has been high. In the present study, we aimed to determine the effects of nonadherence to the IFU on the outcomes.

METHODS

Patients who had undergone EVAR for an infrarenal AAA between 2005 and 2013 were included. IFU nonadherence was defined as any violation of device-specific IFU criteria and was compared with IFU adherence. The primary outcomes were all-cause mortality, aneurysm-related mortality, AAA rupture, graft-related adverse events (GRAEs), including limb-related adverse events, and type Ia endoleaks. A second aim was to study whether the prevalence of EVAR performed outside the IFU has increased over time.

RESULTS

A total of 258 patients were included, 144 (55.8%) of whom had been treated according to the IFU and 114 (44.2%) outside the IFU. In the IFU nonadherence group, all-cause mortality (hazard ratio [HR], 1.39; 95% confidence interval [CI], 1.02-1.89; P = .037) and aneurysm-related mortality (HR, 5.1; 95% CI, 1.4-18.6; P = .015), and the incidence of AAA rupture (HR, 5.4; 95% CI, 1.1-26.6; P = .036) and GRAEs (HR, 1.7; 95% CI, 1.1-2.8; P = .025). No significant association was found between the incidence of type Ia endoleaks and neck-related IFU or limb-related adverse events and iliac-related IFU. However, neck length was a risk factor for type Ia endoleaks (HR, 18.2, 95% CI, 6.3-52.2; P < .001), aneurysm-related mortality (HR, 8.7; 95% CI, 1.8-41.6; P = .007), AAA rupture (HR, 21.7; 95% CI, 2.8-166; P = .003), and GRAEs (HR, 4.4; 95% CI, 2.0-9.7; P < .001). An IFU violation regarding neck angulation was also a risk factor for all-cause mortality (HR, 2.0; 95% CI, 1.1-3.7; P = .032), aneurysm-related mortality (HR, 7.6; 95% CI, 1.4-42.8; P = .021), AAA rupture (HR, 79.4; 95% CI, 6.3-999; P = .001), and GRAEs (HR, 4.3; 95% CI, 1.9-9.5; P < .001). The prevalence of EVAR performed outside the IFU did not increase over time.

CONCLUSIONS

Performing EVAR outside the IFU had a negative effect on outcomes, including all-cause mortality, aneurysm-related mortality, AAA rupture, and GRAEs. Neck angulation and neck length seemed to be the most crucial aneurysm characteristics.

Для цитирования: Hahl Tilda, Protto Sara, Järvenpää Valtteri, Uurto Ilkka, Väärämäki Suvi, Suominen Velipekka. Long-term outcomes of endovascular aneurysm repair according to instructions for use adherence status.. Journal of vascular surgery. 2022. DOI: 10.1016/j.jvs.2022.03.010.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.