DOI 10.1118/1.3644843

Медицинская библиотека · Medical physics · 2011

Dosimetry of a new P-32 ophthalmic applicator.

Choi Chang Heon, Han Hyon Soo, Son Kwang-Jae, Park Ul Jae, Lee Jun Sig, Wee Won Ryang, Ha Sung Whan, Kim Il Han, Ye Sung-Joon

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: The potential of P-32 ophthalmic applicator irradiation after pterygium excision has been demonstrated as an alternative to Sr∕Y-90 irradiation. This study aimed to provide the clinical dosimetry for this new applicator.

METHODS

The prototype of a cylindrical P-32 applicator was fabricated according to the Monte Carlo (MC)-based design study. At a nominal activity of 6 mCi (0.22 GBq), the absorbed dose rate at the front surface (i.e., reference dose rate) was measured by using an extrapolation ionization chamber (EC). The radiochromic film (RCF) was also used to measure the reference dose, axial depth dose distributions and transaxial dose profiles at various depths in water.

RESULTS

The reference dose rate was 3.89 ± 0.14 cGy∕s for EC and 3.84 ± 0.25 cGy∕s for RCF. The depth dose rate was reduced approximately by an order of magnitude for every 2 mm depth in water. Measured depth doses in depths of 0.5-2.5 mm agreed with Monte Carlo data within ±3%. Due to nonuniform absorption of P-32 into an absorbent disk, the dose profiles were not symmetric and decreased more rapidly toward the periphery than those predicted by the MC. The authors confirmed no leakage of P-32 activities and negligible exposure rate around the hand grip of the applicator.

CONCLUSIONS

The P-32 applicator can deliver therapeutic doses to the surface of the conjunctiva, while sparing the lens better than Sr∕Y-90 applicators. The doses at any points from the P-32 applicator could be calculated by using the measured dosimetry data. They also confirmed no leakage of the source, reliable integrity of the applicator, and negligible exposure level around the hand grip of the applicator. However, due to a possibility of nonuniform distributions of P-32 in an absorbent disk, measuring dose profiles as well as the reference dose rate for every new applicator would be recommended.

Для цитирования: Choi Chang Heon, Han Hyon Soo, Son Kwang-Jae, Park Ul Jae, Lee Jun Sig, Wee Won Ryang, Ha Sung Whan, Kim Il Han, Ye Sung-Joon. Dosimetry of a new P-32 ophthalmic applicator.. Medical physics. 2011. DOI: 10.1118/1.3644843.

Предметные рубрики MeSH