In vitro use of standard fluid infusion central venous catheter for slow continuous ultrafiltration feasibility assessment.

Nalesso Federico, Garzotto Francesco, Gobbi Laura, Cattarin Leda, Calò Lorenzo A

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Slow continuous ultrafiltration (SCUF) by central venous catheter (CVC) for hemodialysis is a simple extracorporeal ultrafiltration that can reduce and control fluid overload in patients with chronic or acute chronic heart failure unresponsive to medical therapy. In order to avoid complications and risks related to the catheterization with a standard CVC for hemodialysis to provide the SCUF, considering that hospitalized patients affected by congestive heart failure commonly undergo CVC placement for medications delivering, we tested an in vitro model to use a standard CVC for infusion to perform SCUF.

METHODS

We performed an in vitro SCUF experimental model through two different lumens of a triple (8Fr × 20 cm Tri-lumen catheter Kit-Envans Extra) and a quad-lumen CVC (8.5Fr × 20 cm Quad-lumen catheter kit-Benefis Medical Devices) commonly used in our Intensive Care Unit for fluids and medications infusions. We used Prismaflex with the HF-20 set (Baxter, IL, USA) to perform the SCUF treatment.

RESULTS

Our in vitro data confirm the technical feasibility of the use of standard CVC for fluid infusion to perform a SCUF treatment with a theoretical weight loss of up to 200 ml/h by a blood flow ranging from 30 to 45 ml/min.

CONCLUSIONS

The use of standard infusion CVC could be utilized in the intensive care unit to perform SCUF not exposing patients to the risks and complications related to the placement and permanence of CVC for hemodialysis.

Для цитирования: Nalesso Federico, Garzotto Francesco, Gobbi Laura, Cattarin Leda, Calò Lorenzo A. In vitro use of standard fluid infusion central venous catheter for slow continuous ultrafiltration feasibility assessment.. Artificial organs. 2022. DOI: 10.1111/aor.14253.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.