Comparing HemoCue® and Quantitative Buffy Coat® and Coulter Counter-measured haemoglobin concentrations in African children with acute uncomplicated malaria: a Bland-Altman analysis.

Ayuen Dhol S, Olupot-Olupot Peter, Muhindo Rita, Onyamboko Marie A, Ajayi Seun, Chimjinda Natenapa, Taya Chiraporn, Uyoga Sophie, Williams Thomas N, Maitland Kathryn, Fanello Caterina, Day Nicholas P J, Taylor Walter R, Mukaka Mavuto

СтатьяСравнительное исследованиеРандомизированное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Anaemia is a deleterious consequence of malaria, and its accurate diagnosis is crucial for effective management. However, laboratory methods for measuring haemoglobin (Hb) concentration, like the Coulter Counter and the Quantitative Buffy Coat® (QBC®), are costly and not widely accessible in resource-limited settings. The point-of-care HemoCue® test is a cheaper alternative and suitable in rural areas. The study aimed to determine the level of agreement between Coulter Counter/QBC® vs. HemoCue®-measured Hb concentrations by Bland-Altman analysis.

METHODS

As part of a randomized, placebo-controlled trial of single low-dose primaquine in Ugandan and Congolese children with acute uncomplicated Plasmodium falciparum malaria, Hb concentrations were measured on days 0, 3, 7, and 28 using Coulter Counter (Uganda, n = 1880 paired values), QBC® (DR Congo, n = 1984 paired values) and HemoCue® Hb-301™. The predefined clinically acceptable limits were set at ± 0.5 g/dL.

RESULTS

The Bland-Altman analysis showed that the HemoCue® minus Coulter Counter mean Hb difference was - 0.15 g/dL with lower and upper limits of agreement of - 3.68 g/dL and 3.39 g/dL, respectively. Corresponding HemoCue® minus QBC® values were - 0.23 g/dL, - 1.66 g/dL and 1.22 g/dL. Linear regression of Hb concentration differences vs. mean Hb concentrations showed negative correlations: r = - 0.43 and r = - 0.34 for HemoCue® vs. Coulter Counter and HemoCue® vs. QBC®, respectively.

CONCLUSIONS

Compared to Coulter and QBC®, mean HemoCue® measured Hb concentrations were lower and, compared to the Coulter or QBC® methods, had an overall tendency to measure lower Hb concentrations with increasing Hb concentrations. Upper and lower limits of agreement were wider than the predefined clinically acceptable limits of ± 0.5 g/dL. HemoCue® should be used with caution in settings where decisions about blood transfusions are made.

Для цитирования: Ayuen Dhol S, Olupot-Olupot Peter, Muhindo Rita, Onyamboko Marie A, Ajayi Seun, Chimjinda Natenapa, Taya Chiraporn, Uyoga Sophie, Williams Thomas N, Maitland Kathryn, Fanello Caterina, Day Nicholas P J, Taylor Walter R, Mukaka Mavuto. Comparing HemoCue® and Quantitative Buffy Coat® and Coulter Counter-measured haemoglobin concentrations in African children with acute uncomplicated malaria: a Bland-Altman analysis.. Malaria journal. 2025. DOI: 10.1186/s12936-025-05318-5.