Effectiveness of the Spot Vision Screener using updated 2021 AAPOS guidelines.

Peterseim Mae Millicent W, Trivedi Rupal H, Monahan Savannah Reaves, Smith Samantha M, Bowsher James D, Alex Anastasia, Wilson M Edward, Wolf Bethany J

СтатьяПоддержка NIH: внешняя программаПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the Spot Vision Screener according to updated 2021 AAPOS Vision Screening Committee guidelines for instrument-based pediatric vision screen validation.

METHODS

As part of an IRB-approved ongoing prospective study, children were screened with the Spot prior to a complete examination.

RESULTS

Spot screening was successful in 1,036 of 1,090 children (95%). Forty-eight percent of participants were referred for further screening using the Spot manufacturer guidelines, and 40% of all children were found to have a 2021 amblyopia risk factor or visually significant refractive error by gold standard examination. The Spot recommendation compared reasonably well to the 2021 criteria, with an overall sensitivity of 0.88 and a specificity of 0.78. Applying updated guidelines to the Spot for hyperopia, anisometropia, and astigmatism yielded moderate-to-poor sensitivity (0.27-0.77) but excellent specificity (>0.9). The area under the curve of the receiver operating characteristic analysis demonstrates overall good prediction performance for the Spot for each diagnosis-myopia, hyperopia, astigmatism, anisometropia (range, 0.87-0.97). Results of our study suggest increasing the instrument referral criterion for astigmatism from 1.5 D (manufacturer thresholds of the screener used in this study) to 2 D in older children. Decreasing the anisometropia cut-off from 1 D to 0.75 D would improve sensitivity from 0.59 to >0.8.

CONCLUSIONS

In our study population, the overall predictive ability of the Spot is good, with a sensitivity of 0.88 and a specificity of 0.78. We recommend specific device refractive referral criteria to maximize screening effectiveness using the updated AAPOS guidelines.

Для цитирования: Peterseim Mae Millicent W, Trivedi Rupal H, Monahan Savannah Reaves, Smith Samantha M, Bowsher James D, Alex Anastasia, Wilson M Edward, Wolf Bethany J. Effectiveness of the Spot Vision Screener using updated 2021 AAPOS guidelines.. Journal of AAPOS : the official publication of the American Association for Pediatric Ophthalmology and Strabismus. 2023. DOI: 10.1016/j.jaapos.2022.11.019.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.