MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1186/s12894-024-01458-z

Медицинская библиотека · BMC urology · 2024

Effect of nephrostomy sheath size on renal pelvic pressure during endoscopic combined intrarenal surgery: artificial kidney model study.

Iguchi Takashi, Yamashita Shimpei, Kohjimoto Yasuo, Tanioku Tadashi, Kawamata Tomoyuki, Hara Isao

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

This study aimed to evaluate the intrarenal pelvic pressure in endoscopic combined intrarenal surgery using an artificial kidney model.

METHODS

An artificial kidney model was created using the Urovac evacuator™. Four sizes of nephrostomy sheaths (MIP-L: 25/26 Fr, MIP-M: 16.5/17.5 Fr, MIP-S: 11/12 Fr, MIP-XS: 8.5/9.5 Fr) and two sizes of ureteral access sheaths (12/14 Fr and 10/12 Fr) were installed into the model. For each combination of nephrostomy and ureteral access sheath, renal pelvic pressure was measured with and without insertion of the retrograde flexible ureteroscope. Irrigation from the nephroscope was adjusted to 40-160 mmHg using an automatic irrigation device, and the irrigation of the ureteroscope was by spontaneous dripping at 80 cmH2O. Conditions were measured six times, and the renal pelvic pressure was compared in different conditions.

RESULTS

Without ureteroscope insertion through the ureteral access sheath, the renal pelvic pressure never exceeded 30 mmHg. Meanwhile, when the ureteroscope was inserted, the renal pelvic pressure increased as the nephrostomy sheath and ureteral access sheath became narrower and as the irrigation pressure increased. Intrarenal pelvic pressure exceeded 30 mmHg when the irrigation pressure was increased in 12/14 Fr ureteral access sheath when MIP-XS was used, and in 10/12 Fr ureteral access sheath when MIP-XS and MIP-S were used.

CONCLUSIONS

The use of a thin nephrostomy sheath in endoscopic combined intrarenal surgery can lead to increased intrarenal pelvic pressure. Although our results are from an artificial kidney model, special care is suggested to be required when using a retrograde flexible ureteroscope simultaneously in treatment of patients.

Для цитирования: Iguchi Takashi, Yamashita Shimpei, Kohjimoto Yasuo, Tanioku Tadashi, Kawamata Tomoyuki, Hara Isao. Effect of nephrostomy sheath size on renal pelvic pressure during endoscopic combined intrarenal surgery: artificial kidney model study.. BMC urology. 2024. DOI: 10.1186/s12894-024-01458-z.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.