MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1159/000048305

Медицинская библиотека · Ophthalmologica. Journal international d'ophtalmologie. International journal of ophthalmology. Zeitschrift fur Augenheilkunde · 2002

Typical ultrasound biomicroscopic findings seen in ocular hypotony.

Roters Sigrid, Engels Bert F, Szurman Peter, Krieglstein Günter K

СтатьяПеревод пока не подготовлен

Аннотация

UNLABELLED: Detailed imaging of the ciliary body region by means of ultrasound biomicroscopy can provide information that is crucial for further treatment whereas clinical evaluation and 10-MHz standard B scan are less helpful.

BACKGROUND

To evaluate the suitability of ultrasound biomicroscopy (UBM) for elucidating the causes of ocular hypotony.

PATIENTS AND

METHODS

A retrospective chart review was performed to determine the usefulness of UBM in the evaluation of ocular hypotony. UBM, B scan, and a detailed clinical examination had been performed. Sixty patient charts with prolonged ocular hypotony of different causes were reviewed from January 1994 to December 1998. The information obtained by the UBM was classified into three groups: 'diagnostic' (group 1), 'helpful' (group 2), 'not helpful' (group 3).

RESULTS

Of the 60 eyes, 45 UBM examinations (75%) were in group 1, 9 eyes (15%) were assigned to group 2 and 6 eyes (10%) were in group 3. The duration of hypotony had no influence.

CONCLUSION

In cases of ocular hypotony, the cause was related to ciliary body pathologies in 80%. Only UBM could distinguish tractional from dehiscence ciliary body detachment which required a different management approach. UBM with its high resolution imaging of the anterior segment - including the ciliary body - is highly suitable for diagnostic clarification.

Для цитирования: Roters Sigrid, Engels Bert F, Szurman Peter, Krieglstein Günter K. Typical ultrasound biomicroscopic findings seen in ocular hypotony.. Ophthalmologica. Journal international d'ophtalmologie. International journal of ophthalmology. Zeitschrift fur Augenheilkunde. 2002. DOI: 10.1159/000048305.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.