A case of cardiac strangulation following epicardial pacemaker implantation.

Miyagi Chihiro, Ochiai Yoshie, Ando Yusuke, Hisahara Manabu, Baba Hironori, Takigawa Tomoya, Tokunaga Shigehiko

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

An 8-year-old boy had undergone permanent epicardial pacemaker implantation with a Y-shaped bipolar ventricular lead on day 6 after birth for treatment of congenital complete atrioventricular block. He was found to have pulmonary stenosis and mitral stenosis by follow-up echocardiography. Further studies including computed tomography and cardiac catheterization revealed that the pacemaker lead had completely encircled the cardiac silhouette and was in a state of "cardiac strangulation". We removed the previous pacing leads and generator and implanted a new epicardial dual-chamber pacing system in the right atrium and right ventricle. Additionally, an expanded polytetrafluoroethylene sheet was placed between the new leads and the heart to prevent recurrence of cardiac strangulation.

Для цитирования: Miyagi Chihiro, Ochiai Yoshie, Ando Yusuke, Hisahara Manabu, Baba Hironori, Takigawa Tomoya, Tokunaga Shigehiko. A case of cardiac strangulation following epicardial pacemaker implantation.. General thoracic and cardiovascular surgery. 2020. DOI: 10.1007/s11748-020-01337-y.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.