Медицинская библиотека · Radiology · 1997

Patient dose due to colon examination: dose assessment and results from a survey in The Netherlands.

Geleijns J, Broerse J J, Shaw M P, Schultz F W, Teeuwisse W, van Unnik J G, Zoetelief J

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To determine the effective dose to the patient during radiographic colon examination.

MATERIALS AND

METHODS

The integral dose-area product was measured during colon examination in 1,733 patients aged 18-94 years. The effective dose was estimated from the dose-area product through computer simulations of radiation transport in anthropomorphic phantoms. The relation between patient dose and imaging or radiographic technique was considered. Patient dose from a biphasic colon examination was compared to that from a double-contrast examination.

RESULTS

The factors for converting dose-area product to effective dose were 0.29 mSv x Gy(-1) x cm(-2) and 0.27 mSv x Gy(-1) x cm(-2) for the biphasic and the double-contrast studies, respectively. The average dose-area product for the biphasic colon examination was 21 Gy x cm2, of which 13 Gy x cm2 was attributed to the double-contrast views. The average dose-area product was 29 Gy x cm2 (range, 18-53 Gy x cm2); the average effective dose was 4.7 mSv (range, 2.7-8.4 mSv).

CONCLUSION

Careful selection of the radiologic technique resulted in a surprisingly low dose during the biphasic colon examination. It is recommended that additional filtration of at least 0.1-mm copper be applied and that a screen-film combination with a speed class of at least 400 be used. Dose reduction when using digital techniques is often not realized in clinical practice.

Для цитирования: Geleijns J, Broerse J J, Shaw M P, Schultz F W, Teeuwisse W, van Unnik J G, Zoetelief J. Patient dose due to colon examination: dose assessment and results from a survey in The Netherlands.. Radiology. 1997. DOI: 10.1148/radiology.204.2.9240552.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.