MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1038/s41433-022-02070-x

Медицинская библиотека · Eye (London, England) · 2022

Diagnostic accuracy of different keratoconus detection indices of pentacam in paediatric eyes.

Hashem Ahmed Osama, Aziz Bassem Fayez, Wahba Sherine Shafik, Roshdy Maged Maher, Elawamry Amr Ismail

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

No diagnostic gold standard for keratoconus in children and adolescents exists. Our objective was investigating the diagnostic accuracy of various indices for keratoconus (KC) detection in paediatric eyes.

METHODS

All retrievable data of significance from 432 normal right paediatric eyes and 48 eyes of paediatric KC and forme fruste KC (FFKC), imaged by use of a rotating Scheimpflug camera (Oculyzer II, Pentacam HR) between December 2013 and October 2018 at Watany Eye Hospitals, Cairo, Egypt, including Scheimpflug images data, were collected. The area under the receiver operating characteristic curve (AUROC) was calculated for different indices in this retrospective descriptive study.

RESULTS

All 36 tested indices showed discriminative power differentiating KC and FFKC from normal corneas (AUROC P-value <0.05), except AC volume, AC angle, and horizontal decentrations of the steepest and thinnest points. The 32 indices showed variable degrees of diagnostic accuracy. The highest AUROC was that of the corneal assessment index from the relational thickness and other OCULUS values (CAIRO 8). Only 8 indices showed non-inferiority to it, namely, Ambrosio's relational thickness maximum (ART max) and avg, the pachymetric progression index maximum (PPI max) and avg, the back elevation from the best-fit toric ellipsoid (BE BFTE), the KC index (KI), the topographic KC indices (TKC), and the index of height decentration (IHD) (P > 0.05).

CONCLUSIONS

The 8 most useful rotating Scheimpflug imaging indices for KC detection in paediatric eyes are CAIRO 8 followed by ART max and avg, PPI max and avg, BE BFTE, KI, TKC, and IHD.

Для цитирования: Hashem Ahmed Osama, Aziz Bassem Fayez, Wahba Sherine Shafik, Roshdy Maged Maher, Elawamry Amr Ismail. Diagnostic accuracy of different keratoconus detection indices of pentacam in paediatric eyes.. Eye (London, England). 2022. DOI: 10.1038/s41433-022-02070-x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.