A New Approach to Non-Invasive Microcirculation Monitoring: Quantifying Capillary Refill Time Using Oximetric Pulse Waves.
Аннотация
(1)
Background
To develop a novel capillary refill time measurement system and evaluate its reliability and reproducibility. (2)
Methods
Firstly, the utilization of electromagnetic pressure technology facilitates the automatic compression and instantaneous release of the finger. Secondly, the employment of pressure sensing technology and photoelectric volumetric pulse wave analysis technology enables the dynamic monitoring of blood flow in distal tissues. Thirdly, the subjects were recruited to compare the average measurement time and the number of measurements required for successful measurements. The satisfaction of doctors and patients with the instrument was investigated through the administration of questionnaires. Finally, 71 subjects were recruited and divided into two groups, A and B. Three doctors repeated the measurement of the right index fingers of the subjects. In Group A, the same measuring instrument was used, and the consistency of the measurements was evaluated using the intragroup correlation coefficient. In Group B, one doctor repeated the measurement of each subject three times using the same measuring instrument, and the reproducibility of the CRT was evaluated using the analysis of variance of the repeated measurement data. (3)
Results
The development of the capillary refill time meter was successful, with an average measurement time of 18 s and a single measurement. This study found that doctor-patient satisfaction levels were 98.3% and 100%, respectively. The intraclass correlation coefficient was 0.995 in Group A, and the p-value was greater than 0.05 in Group B. (4)
Conclusions
The non-invasive monitoring of microcirculation has been rendered both rapid and effective, thus paving the way for the further mechanization and standardization of this process. The CRT, when measured using the capillary refill time meter test machine, demonstrated consistent and reproducible results, both when assessed by different researchers and when evaluated across varying measurement sets.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.