MedElevator Войти в исследования
← Поиск публикаций
Медицинская библиотека · The Journal of heart valve disease · 2005

Predicting ATS Open Pivot heart valve performance with computational fluid dynamics.

Dumont Kris, Vierendeels Jan A M, Segers Patrick, Van Nooten Guido J, Verdonck Pascal R

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND AND AIM OF THE STUDY: In-vitro studies on the ATS heart valve have indicated that valve opening is less in an expanding conduit than in a straight conduit.

METHODS

Bileaflet valve behavior was studied using a new computational fluid-structure interaction model. A three-dimensional model of the ATS valve was studied in two geometries, simulating the valve in a geometry with sudden expansion downstream of the valve, and in a straight conduit. Mitral and aortic flow patterns were simulated.

RESULTS

The ATS valve in the expanding geometry showed opening to a maximum angle of 77.5 degrees; this was confirmed in previous clinical and in-vitro studies. The mean and maximum transvalvular Doppler pressure gradients were 1.1 and 4.3 mmHg, respectively. The maximum shear stress calculated on the leaflet was 25 Pa. Maximum opening of the valve was achieved in the straight conduit; with mean and maximum pressure gradients of 2.1 and 4.6 mmHg, respectively. The maximum shear stress calculated on the leaflet was 35 Pa.

CONCLUSION

The results of this numerical study confirmed that valve hemodynamics and leaflet motion were dependent on the geometrical conditions of the valve: the presence of a diverging flow influenced the maximum opening angle of the valve leaflets. This model could be used to predict pressure gradients, effective orifice area, performance index and shear stress loading of mechanical heart valves, and in future will serve as a major research tool to characterize the hemodynamics of existing and new mechanical heart valves.

Для цитирования: Dumont Kris, Vierendeels Jan A M, Segers Patrick, Van Nooten Guido J, Verdonck Pascal R. Predicting ATS Open Pivot heart valve performance with computational fluid dynamics.. The Journal of heart valve disease. 2005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.