DOI 10.1093/bja/aet282

Медицинская библиотека · British journal of anaesthesia · 2013

Bioreactance is not reliable for estimating cardiac output and the effects of passive leg raising in critically ill patients.

Kupersztych-Hagege E, Teboul J-L, Artigas A, Talbot A, Sabatier C, Richard C, Monnet X

Оценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Bioreactance estimates cardiac output in a non-invasive way. We evaluated the ability of a bioreactance device (NICOM®) to estimate cardiac index (CI) and to track relative changes induced by volume expansion.

METHODS

In 48 critically ill patients, we measured CI estimated by the NICOM® device (CINicom) and by transpulmonary thermodilution (CItd, PiCCO2™ device) before and after a 500 ml saline infusion. Before volume expansion, we performed a passive leg raising (PLR) test and measured the changes it induced in CINicom and in pulse contour analysis-derived CI.

RESULTS

Considering the values recorded before PLR and before and after volume expansion (n=144), the bias (lower and upper limits of agreement) between CItd and CINicom was 0.9 (-2.2 to 4.1) litre min(-1) m(-2). The percentage error was 82%. There was no significant correlation between the changes in CItd and CINicom induced by volume expansion (P=0.24). An increase in CI estimated by pulse contour analysis >9% during the PLR test predicted fluid responsiveness with a sensitivity of 84% (95% confidence interval 60-97%) and a specificity of 97% (95% confidence interval 82-100%). The area under the receiver operating characteristic curve constructed to test the ability of the PLR-induced changes in CINicom in predicting fluid responsiveness did not differ significantly from 0.5 (P=0.77).

CONCLUSIONS

The NICOM® device cannot accurately estimate the cardiac output in critically ill patients. Moreover, it could not predict fluid responsiveness through the PLR test.

Для цитирования: Kupersztych-Hagege E, Teboul J-L, Artigas A, Talbot A, Sabatier C, Richard C, Monnet X. Bioreactance is not reliable for estimating cardiac output and the effects of passive leg raising in critically ill patients.. British journal of anaesthesia. 2013. DOI: 10.1093/bja/aet282.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.