MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1080/14992027.2018.1479783

Медицинская библиотека · International journal of audiology · 2018

Successful application and timing of a remote network for intraoperative objective measurements during cochlear implantation surgery.

Yanov Yuri, Kuzovkov Vladislav, Sugarova Serafima, Levin Sergey, Lilenko Andrei, Kliachko Dmitrii

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To determine the time needed to remotely perform a set of intraoperative measurements during cochlear implantation surgery and to compare it to the time needed to perform the same measurements in theatre.

DESIGN

Prospective two-arm study comparing a local with a remote measurement setting. Three intraoperative measurements (Impedance Field Telemetry (IFT), evoked compound action potential (ECAP) and evoked stapedius reflex test (eSRT)) were performed with the audiologist present in the operating theatre (i.e. locally) or with the audiologist in his/her office (i.e. remotely). The time needed to complete the measurements, in total and individually, were measured and compared.

STUDY SAMPLE: Fifty cochlear implant recipients, aged 0.7-48 years

Results

IFT, ECAP and eSRT were performed successfully in all participants. Comparing locally and remotely performed measurements, IFT did not differ significantly, ECAP threshold or slope did not differ significantly differ in any single channel and eSRT measurements did not differ significantly except for in one of six channels. Remote measurements took 8.6 min less to perform than did local measurements (10.04 vs. 18.64 min); a significant difference (p < 0.001).

CONCLUSIONS

Using a remote network connection for intraoperative objective measurements is an efficient and safe way to perform measurements during cochlear implantation surgery.

Для цитирования: Yanov Yuri, Kuzovkov Vladislav, Sugarova Serafima, Levin Sergey, Lilenko Andrei, Kliachko Dmitrii. Successful application and timing of a remote network for intraoperative objective measurements during cochlear implantation surgery.. International journal of audiology. 2018. DOI: 10.1080/14992027.2018.1479783.