DOI 10.1080/14670100.2017.1359457

Медицинская библиотека · Cochlear implants international · 2017

Effects of electrode deactivation on speech recognition in multichannel cochlear implant recipients.

Schvartz-Leyzac Kara C, Zwolan Teresa A, Pfingst Bryan E

Оценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

The objective of the current study is to evaluate how speech recognition performance is affected by the number of active electrodes that are turned off in multichannel cochlear implants. Several recent studies have demonstrated positive effects of deactivating stimulation sites based on an objective measure in cochlear implant processing strategies. Previous studies using an analysis of variance have shown that, on average, cochlear implant listeners' performance does not improve beyond eight active electrodes. We hypothesized that using a generalized linear mixed model would allow for better examination of this question.

METHODS

Seven peri- and post-lingual adult cochlear implant users (eight ears) were tested on speech recognition tasks using experimental MAPs which contained either 8, 12, 16 or 20 active electrodes. Speech recognition tests included CUNY sentences in speech-shaped noise, TIMIT sentences in quiet as well as vowel (CVC) and consonant (CV) stimuli presented in quiet and in signal-to-noise ratios of 0 and +10 dB.

RESULTS

The speech recognition threshold in noise (dB SNR) significantly worsened by approximately 2 dB on average as the number of active electrodes was decreased from 20 to 8. Likewise, sentence recognition scores in quiet significantly decreased by an average of approximately 12%.

DISCUSSION/

CONCLUSION

Cochlear implant recipients can utilize and benefit from using more than eight spectral channels when listening to complex sentences or sentences in background noise. The results of the current study suggest a conservative approach for turning off stimulation sites is best when using site-selection procedures.

Для цитирования: Schvartz-Leyzac Kara C, Zwolan Teresa A, Pfingst Bryan E. Effects of electrode deactivation on speech recognition in multichannel cochlear implant recipients.. Cochlear implants international. 2017. DOI: 10.1080/14670100.2017.1359457.

Предметные рубрики MeSH