Repeatability and agreement of ARK-30 autorefraction after cataract surgery.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

de Juan Victoria, Herreras Jose Maria, Martin Raul, Morejon Angela, Perez Inmaculada, Rio-Cristobal Ana, Rodriguez Guadalupe

Аннотация

BACKGROUND: To evaluate the intra-test variability of ARK-30 handheld autorefractor and the agreement with subjective refraction and retinoscopy after uneventful cataract surgery.

DESIGN: Prospective and non-randomized study that included 6 visits by patients undergoing uneventful cataract surgery at IOBA (Instituto de Oftalmobiología Aplicada) Eye Institute (University of Valladolid).

PARTICIPANTS: The mean age of the 79 patients was 66.5years (range 23-90years). For the 124 eyes, the mean spherical equivalent of the sample at baseline visit was -3.59±6.28D (range -21.00D to +4.44D).

METHODS: Automated refraction was performed on follow-up visits 1day and weekly for 4weeks. Retinoscopy and subjective refraction were conducted at the Week 4 follow up.

MAIN OUTCOME MEASURES: Automated refraction.

RESULTS: Sphere, cylinder and mean spherical equivalent, J(0) and J(45) coefficient variabilities were low in all visits. Standard deviations and the limits of agreement were smallest for the last visit. Subjective refraction sphere and cylinder values were more positive than autorefraction by 0.12±0.53D (P=0.031) and 0.23±0.42D (P<0.001), respectively. Comparison between autorefraction and retinoscopy showed a similar trend with the sphere and cylinder differences, 0.32±0.77D and 0.38±0.43D (P<0.05), respectively.

CONCLUSIONS: The ARK-30 is sufficiently accurate and repeatable for automated refraction after uneventful cataract surgery. This instrument may be useful for monitoring refractive outcome in these patients.