Postoperative displacement of deep brain stimulation electrodes related to lead-anchoring technique.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Contarino M Fiorella, Bot Maarten, Speelman Johannes D, de Bie Rob M A, Tijssen Marina A, Denys Damiaan, Bour Lo J, Schuurman P Richard, van den Munckhof Pepijn

Аннотация

BACKGROUND: Displacement of deep brain stimulation (DBS) electrodes may occur after surgery, especially due to large subdural air collections, but other factors might contribute.

OBJECTIVE: To investigate factors potentially contributing to postoperative electrode displacement, in particular, different lead-anchoring techniques.

METHODS: We retrospectively analyzed 55 patients (106 electrodes) with Parkinson disease, dystonia, tremor, and obsessive-compulsive disorder in whom early postoperative and long-term follow-up computed tomography (CT) was performed. Electrodes were anchored with a titanium microplate or with a commercially available plastic cap system. Two independent examiners determined the stereotactic coordinates of the deepest DBS contact on early postoperative and long-term follow-up CT. The influence of age, surgery duration, subdural air volume, use of microrecordings, fixation method, follow-up time, and side operated on first was assessed.

RESULTS: Subdural air collections measured on average 4.3 ± 6.2 cm. Three-dimensional (3-D) electrode displacement and displacement in the X, Y, and Z axes significantly correlated only with the anchoring method, with larger displacement for microplate-anchored electrodes. The average 3-D displacement for microplate-anchored electrodes was 2.3 ± 2.0 mm vs 1.5 ± 0.6 mm for electrodes anchored with the plastic cap (P = .030). Fifty percent of the microplate-anchored electrodes showed 2-mm or greater (potentially relevant) 3-D displacement vs only 25% of the plastic cap-anchored electrodes (P < .01).

CONCLUSION: The commercially available plastic cap system is more efficient in preventing postoperative DBS electrode displacement than titanium microplates. A reliability analysis of the electrode fixation is warranted when alternative anchoring methods are used.