MedElevator Войти в исследования
← Поиск публикаций

DOI 10.4187/respcare.05703

Медицинская библиотека · Respiratory care · 2018

Imposed Work of Breathing During High-Frequency Oscillatory Ventilation in Spontaneously Breathing Neonatal and Pediatric Models.

Bordessoule Alice, Piquilloud Lise, Lyazidi Aissam, Moreira Amelia, Rimensberger Peter C

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

High-frequency oscillatory ventilation (HFOV) is used in cases of neonatal and pediatric acute respiratory failure, sometimes even as the primary ventilatory mode. Allowing patients (at least neonates) on HFOV to breathe spontaneously soon after intubation has been shown to be feasible, and this is becoming a more generally used approach for infants and small children. However, such an approach may increase the imposed work of breathing (WOB), raising the question of whether the imposed WOB varies with the use of newer-generation HFOV devices, which operate according to different functional principles.

METHODS

A bench test was designed to compare the pressure-time product (PTP), a surrogate marker of the imposed WOB, produced with the use of 7 HFOV devices. Scenarios corresponding to various age groups (preterm newborn [1 kg], term newborn [3.5 kg], infant [10 kg], and child [25 kg]) and 2 respiratory system conditions (physiologic and pathologic) were tested.

RESULTS

The PTP varied between devices and increased with the oscillation frequency for all devices, independent of the respiratory system condition. Furthermore, the PTP increased with age and was higher for physiologic than for pathologic respiratory system conditions. We considered a change of ≥ 20% as being of clinically relevant; the effect of oscillation frequency was the most important parameter influencing imposed WOB during spontaneous breathing.

CONCLUSIONS

Variations in imposed WOB, as expressed by PTP values, during spontaneous breathing depend mainly on the oscillator frequency, respiratory system condition, and, though to a lesser extent, on the device itself.

Для цитирования: Bordessoule Alice, Piquilloud Lise, Lyazidi Aissam, Moreira Amelia, Rimensberger Peter C. Imposed Work of Breathing During High-Frequency Oscillatory Ventilation in Spontaneously Breathing Neonatal and Pediatric Models.. Respiratory care. 2018. DOI: 10.4187/respcare.05703.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.