Pulmonary vein isolation using the Rhythmia mapping system: Verification of intracardiac signals using the Orion mini-basket catheter.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Anter Elad, Tschabrunn Cory M, Contreras-Valdes Fernando M, Li Jianqing, Josephson Mark E
Аннотация
BACKGROUND: During pulmonary vein isolation (PVI), a circular lasso catheter is positioned at the junction between the left atrium (LA) and the pulmonary vein (PV) to confirm PVI. The Rhythmia mapping system uses the Orion mini-basket catheter with 64 electrodes instead of the lasso catheter. However, its feasibility to determine PVI has not been studied.
OBJECTIVE: The purpose of this study was to compare signals between the mini-basket and lasso catheters at the LA-PV junction.
METHODS: In 12 patients undergoing PVI using Rhythmia, the mini-basket and lasso catheters were placed simultaneously at the LA-PV junction for baseline and post-PVI signal assessment. Pacing from both catheters was performed to examine the presence of exit block.
RESULTS: At baseline, recordings of LA and PV potentials were concordant in all PVs. However, after PVI, concordance between the catheters was only 68%. Discordance in all cases resulted from loss of PV potentials on the lasso catheter with persistence of PV potentials on the mini-basket catheter. In 9 of 13 PVs (69%), these potentials represented true PV potentials that were exclusively recorded with the smaller and closely spaced mini-basket electrodes. In the other 4 PVs (31%), these potentials originated from neighboring structures and resulted in underestimation of PVI.
CONCLUSION: The use of the mini-basket catheter alone is sufficient to determine PVI. While it improves recording of PV potentials after incomplete ablation, it is also associated with frequent recording of "PV-like" potentials originating from neighboring structures. In these cases, pacing maneuvers are helpful to determine PVI and avoid excessive ablation.
OBJECTIVE: The purpose of this study was to compare signals between the mini-basket and lasso catheters at the LA-PV junction.
METHODS: In 12 patients undergoing PVI using Rhythmia, the mini-basket and lasso catheters were placed simultaneously at the LA-PV junction for baseline and post-PVI signal assessment. Pacing from both catheters was performed to examine the presence of exit block.
RESULTS: At baseline, recordings of LA and PV potentials were concordant in all PVs. However, after PVI, concordance between the catheters was only 68%. Discordance in all cases resulted from loss of PV potentials on the lasso catheter with persistence of PV potentials on the mini-basket catheter. In 9 of 13 PVs (69%), these potentials represented true PV potentials that were exclusively recorded with the smaller and closely spaced mini-basket electrodes. In the other 4 PVs (31%), these potentials originated from neighboring structures and resulted in underestimation of PVI.
CONCLUSION: The use of the mini-basket catheter alone is sufficient to determine PVI. While it improves recording of PV potentials after incomplete ablation, it is also associated with frequent recording of "PV-like" potentials originating from neighboring structures. In these cases, pacing maneuvers are helpful to determine PVI and avoid excessive ablation.