Pulmonary vein isolation using the Rhythmia mapping system: Verification of intracardiac signals using the Orion mini-basket catheter.

Anter Elad, Tschabrunn Cory M, Contreras-Valdes Fernando M, Li Jianqing, Josephson Mark E

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

During pulmonary vein isolation (PVI), a circular lasso catheter is positioned at the junction between the left atrium (LA) and the pulmonary vein (PV) to confirm PVI. The Rhythmia mapping system uses the Orion mini-basket catheter with 64 electrodes instead of the lasso catheter. However, its feasibility to determine PVI has not been studied.

OBJECTIVE

The purpose of this study was to compare signals between the mini-basket and lasso catheters at the LA-PV junction.

METHODS

In 12 patients undergoing PVI using Rhythmia, the mini-basket and lasso catheters were placed simultaneously at the LA-PV junction for baseline and post-PVI signal assessment. Pacing from both catheters was performed to examine the presence of exit block.

RESULTS

At baseline, recordings of LA and PV potentials were concordant in all PVs. However, after PVI, concordance between the catheters was only 68%. Discordance in all cases resulted from loss of PV potentials on the lasso catheter with persistence of PV potentials on the mini-basket catheter. In 9 of 13 PVs (69%), these potentials represented true PV potentials that were exclusively recorded with the smaller and closely spaced mini-basket electrodes. In the other 4 PVs (31%), these potentials originated from neighboring structures and resulted in underestimation of PVI.

CONCLUSION

The use of the mini-basket catheter alone is sufficient to determine PVI. While it improves recording of PV potentials after incomplete ablation, it is also associated with frequent recording of "PV-like" potentials originating from neighboring structures. In these cases, pacing maneuvers are helpful to determine PVI and avoid excessive ablation.

Для цитирования: Anter Elad, Tschabrunn Cory M, Contreras-Valdes Fernando M, Li Jianqing, Josephson Mark E. Pulmonary vein isolation using the Rhythmia mapping system: Verification of intracardiac signals using the Orion mini-basket catheter.. Heart rhythm. 2015. DOI: 10.1016/j.hrthm.2015.05.019.