DOI 10.1016/s0003-4975(98)01102-3

Медицинская библиотека · The Annals of thoracic surgery · 1998

Infrared spectroscopy in the evaluation of the process of calcification of valvular bioprostheses.

Shen M, Lajos P S, Farge D, Daudon M, Carpentier S M, Chen L, Martinet B, Carpentier A F

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

The methods currently used to analyze the process of calcification of bioprostheses give only global information on calcium deposition. We investigated the potential advantage of infrared spectroscopy, which makes it possible to analyze the various components of the calcification process, ie, lipids, proteins, and calcium deposits.

METHODS

Sixty porcine aortic leaflets were fixed in 0.6% glutaraldehyde and then subsequently implanted in 10-day-old Wistar rats. The valve leaflets were removed 2, 7, 14, 21, 35, and 56 days after implantation.

RESULTS

Before implantation infrared spectroscopic analysis revealed the presence of proteins only. On day 2 after implantation, all valves showed minor lipid deposits. On day 7, amorphous calcium phosphate was detected. Between days 7 and 14, crystalline forms of calcium phosphate appeared and amorphous calcium phosphate progressively changed into carbapatite over the 56-day period.

CONCLUSIONS

Infrared spectroscopy yields valuable additional information on the nature and kinetics of the various components of glutaraldehyde-treated tissues after implantation. It may prove to be important in the evaluation of new techniques of calcium mitigation.

Для цитирования: Shen M, Lajos P S, Farge D, Daudon M, Carpentier S M, Chen L, Martinet B, Carpentier A F. Infrared spectroscopy in the evaluation of the process of calcification of valvular bioprostheses.. The Annals of thoracic surgery. 1998. DOI: 10.1016/s0003-4975(98)01102-3.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.