Safety and reliability of the insertable Reveal XT recorder in patients undergoing 3 Tesla brain magnetic resonance imaging.

СтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Haeusler Karl Georg, Koch Lydia, Ueberreiter Juliane, Coban Nalan, Safak Erdal, Kunze Claudia, Villringer Kersten, Endres Matthias, Schultheiss Heinz-Peter, Fiebach Jochen B, Schirdewan Alexander

Аннотация

BACKGROUND: Up to now there is little evidence about the safety and reliability of insertable cardiac monitors (ICMs) in patients undergoing magnetic resonance imaging (MRI).

OBJECTIVE: The purpose of this prospective single-center study (MACPAF; clinicaltrials.govNCT01061931), which we are currently performing, was to evaluate these issues for the ICM Reveal XT at a 3 Tesla MRI scanner in patients undergoing serial brain MRI.

METHODS: We present an interim analysis including 62 brain MRI examinations in 24 patients with paroxysmal atrial fibrillation bearing the Reveal XT. All patients were interviewed for potential ICM-associated clinical symptoms during and after MRI examination. According to the study protocol, data from the Reveal XT were transmitted before and after the MRI examination.

RESULTS: All patients were clinically asymptomatic during the MRI procedure. Moreover, the reliability (ability to detect signals, battery status) of the Reveal XT was unaffected, except for one MRI-induced artifact that was recorded by the ICM, mimicking a narrow complex tachycardia, as similarly recorded in a further study patient bearing the forerunner ICM Reveal DX. No loss of ICM data was observed after the MRI examination.

CONCLUSIONS: The 3 Tesla brain MRI scanning is safe for patients bearing the ICM Reveal XT and does not alloy reliability of the Reveal XT itself. MRI-induced artifacts occur rarely but have to be taken into account.