DOI 10.1016/j.jtha.2023.04.031

Медицинская библиотека · Journal of thrombosis and haemostasis : JTH · 2023

Sonorheometry versus rotational thromboelastometry in trauma: a comparison of diagnostic and prognostic performance.

Rossetto Andrea, Wohlgemut Jared M, Brohi Karim, Davenport Ross

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Rotational thromboelastometry (ROTEM) is used to rapidly identify trauma-induced coagulopathy (TIC) and direct targeted interventions in hemorrhaging trauma patients. A novel technology, Quantra System (HemoSonics), utilizes sonic estimation of elasticity via resonance sonorheometry, avoids mechanical clot interference, and may increase diagnostic accuracy, but there are limited data on bleeding in major trauma patients.

OBJECTIVES

To compare the performance of Quantra with that of ROTEM for rapid diagnosis of TIC and prediction of transfusion requirements and mortality.

METHODS

Samples were collected from adult trauma patients enrolled in a perpetual cohort study upon admission to a single level 1 trauma center between 2020 and 2021. Samples were analyzed using Quantra, ROTEM, multiple electrode aggregometry, and conventional coagulation assays.

RESULTS

Samples from 209 patients were analyzed. Correlations were strong between Quantra and ROTEM parameters (for all, p < .001): clot stiffness (CS) and tissue factor-activated ROTEM (EXTEM) clot amplitude at 5 minutes (A5) (r = 0.90); fibrinogen contribution to CS and tissue factor-activated ROTEM with cytochalasin D (FIBTEM) A5 (r = 0.85); and platelet contribution to CS and EXTEM-FIBTEM A5 (r = 0.73). Although CS showed higher discrimination than EXTEM A5 in detecting TIC (international normalized ratio, >1.2; area under the receiver operating characteristic curve, 0.83 vs 0.79; p = .038), the ability of fibrinogen contribution to CS to detect hypofibrinogenemia (a fibrinogen level of <2g/L) was good but lower than that of FIBTEM A5 (area under the receiver operating characteristic curve, 0.79 vs 0.84; p = .027). There was no difference between Quantra and ROTEM in detecting a platelet count of <150 × 109/L, predicting rapid transfusion or mortality at 6 hours.

CONCLUSION

Quantra and ROTEM have similar diagnostic performances in evaluating TIC and predicting clinically relevant outcomes. Larger studies are required to determine the utility of Quantra for goal-directed treatment of TIC.

Для цитирования: Rossetto Andrea, Wohlgemut Jared M, Brohi Karim, Davenport Ross. Sonorheometry versus rotational thromboelastometry in trauma: a comparison of diagnostic and prognostic performance.. Journal of thrombosis and haemostasis : JTH. 2023. DOI: 10.1016/j.jtha.2023.04.031.