DOI 10.1016/j.cca.2024.119970

Медицинская библиотека · Clinica chimica acta; international journal of clinical chemistry · 2024

Clinical performance and application of novel serum collection "Ser-Col" device in the practice of laboratory diagnosis of infection diseases and several other immunochemical tests.

Harutyunyan Hayk, Hovhannisyan Alvard, Torosyan Hamlet, Hakobjanyan Gohar, Grigoryan Ani, Petrosyan Gayane, Movsisyan Mariam, Poland Dennis, Monkel Richard, Lommen Joan, Tuaillon Edouard, Yenkoyan Konstantin

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Dried blood collection devices might be beneficial for diagnosing infectious diseases in areas far from the medical facilities and in lockdown situations. There are several reports on the efficacy of such applications for qualitative tests. Here we demonstrated the feasibility of a novel Ser-Col blood collection device as a standardized approach for qualitative and quantitative detection of infectious markers and several over immunochemical tests.

METHODS

In the current study, we included 395 adult participants, 191 men and 204 women, with a median age of 41 years, as well as 75 children with a median age of 3 years. Serological status was determined by testing serum samples for three groups of infection diseases: hepatitis A and C, SARS-CoV-2, and herpes family viruses, as well as for thyroid peroxidase (TPO), prolactin, vitamin B12, and folate. Blood collected on the Ser-Col device (Labonovum) was eluted using an automated system (SCAUT Ser-Col automation, Blok System Supply) and manually. Ser-Col results were compared with serum sampled via standard venipuncture considered as the reference.

RESULTS

High correlation coefficients (r = 0.95-0.99) were observed between serum samples collected with Ser-Col and via standard venipuncture for the following tests: anti-HCV, anti-SARS-CoV-2 IgG, anti-HSV-2 IgG, and anti-CMV IgM. Correlation coefficients between Ser-Col and standard venipuncture serum for anti-HSV-1 IgG, anti-CMV IgG, and anti-EBV tests were relatively low (r = 0.73-0.77). Correlation coefficients for anti-TPO, prolactin, vitamin B12, and folate were also characterized with high values (r = 0.97-0.99).

CONCLUSIONS

High accuracy and quantitative correlation were demonstrated between Ser-Col and samples collected by standard venipuncture. Hence, the Ser-Col blood collection device should be considered as a promising alternative for blood collection, storage, and transportation in both adult and pediatric populations.

Для цитирования: Harutyunyan Hayk, Hovhannisyan Alvard, Torosyan Hamlet, Hakobjanyan Gohar, Grigoryan Ani, Petrosyan Gayane, Movsisyan Mariam, Poland Dennis, Monkel Richard, Lommen Joan, Tuaillon Edouard, Yenkoyan Konstantin. Clinical performance and application of novel serum collection "Ser-Col" device in the practice of laboratory diagnosis of infection diseases and several other immunochemical tests.. Clinica chimica acta; international journal of clinical chemistry. 2024. DOI: 10.1016/j.cca.2024.119970.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.